Calibration In General; Notes; General Information; General Procedure For Calibration - JUMO AQUIS touch P Installation Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch P:
Table of Contents

Advertisement

11.1

Notes

11.2

General information

11.2.1

General procedure for calibration

True calibration (calibrating with routines)
WARNING!
During the calibration, the relays and analog output signals assume the config-
ured states!
The actual electrical characteristics of analysis sensors always deviate some-
what from the nominal specifications. The reasons for this include:
Like every measuring instrument, analysis sensors always have a certain
measurement uncertainty that results from manufacturing tolerances.
During use, analysis sensors are exposed to chemical processes. Deposits
and wear phenomena caused by these processes result in changes of the
electrical characteristics of sensors.
To optimize the accuracy of measurements, analysis sensors must be calibrat-
ed. Calibrations are required:
during installation or when changing a sensor
regularly at time intervals that must be specified by the user
if implausible measuring values are displayed
if process conditions change, e. g. as the result of equipment modification
Calibration timers can be configured to provide a regular reminder of when cal-
ibrations are due.
v Chapter 9.10 "Calibration timers", Page 115
Each successfully completed calibration is recorded in the calibration
logbook.
v Chapter 11.3 "Calibration logbook", Page 17
After opening one of the calibration routines in the device, the user is guided
through a process with measurements and entries. In the course of this process,
the calibration values are determined and saved automatically.
Suitable calibration routines are available for every type of analysis sensor. The
individual calibration routines for the various sensor types are described in sep-
arate chapters.
v Chapter 12 "Calibrating a pH measuring chain", Page 21 to Chapter 16 "Cal-
ibrating Ci conductivity sensors", Page 43
Execution of calibration routines requires that the following prerequisites be met:

11 Calibration in general

15

Advertisement

Table of Contents
loading

This manual is also suitable for:

202580

Table of Contents