17 Calibrating universal inputs
Parameter
Temperature com-
pensation
pH compensation
source
56
free chlorine, ph/Temp.-compensated
In the universal inputs calibration default settings for the "free chlorine, ph/T-
compensated" operation mode, the slope calibration is enabled and pre-config-
ured as the only available calibration routine.
Sample screen:
Universal input calibration
default settings in the
"free chlorine, pH/T-com-
pensated" operation
mode
Additional calibration default settings for calibration of chlorine sensors are ex-
plained in the following table.
Possible settings
Selection from Analog selection
Selection from Analog selection
NOTE!
The "Calibration Default Settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in.
v Chapter 8.2.1 "Log-on/Log-out", Page 74
Explanation
Temperature input for automatic sens-
ing of the test/sample solution tempera-
ture during the calibration
pH measuring input for automatic sens-
ing of the test/sample solution pH value
during the calibration