17.2.2
Universal inputs calibration default settings
Calibration default settings for the individual operation modes
The calibration default setting available for selection depends on the configura-
tion settings of the universal input.
Open the calibration default settings:
Device Menu r Calibration r Select Universal Input r
Calibration Default Settings
•
linear scaling
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
The "linear scaling" operation mode permits adjustment of the zero point, slope,
and two-point calibration.
Sample screen:
Universal input calibra-
tion default settings in
the "linear scaling"
operation mode
•
pH value measurement
The calibration default settings correspond to those for the zero-point and two-
point calibration for pH/Redox/NH
v „pH calibration default settings", Page 23
•
Conductivity measurement
The calibration default settings correspond to those for the Ci calibration.
v Chapter 16.2.2 "Calibration default settings for Ci conductivity sensors", Page
44
17 Calibrating universal inputs
analysis inputs.
3
55