Step
Action
6
Insert the optional board in the selected slot. Ensure that the board
is seated correctly.
7
Fill all empty slots with plastic board frames.
8
Re-insert the rear panel of the case and fasten with 3 screws. Make
sure that all the toothed lock washers for securing the screws are
re-inserted (see Step 7). The screws represent the protection con-
ductor connection to the rear panel of the case.
9
Reinsert all the plug-in screw terminals and interface cables that
were removed in Step 4 into the correct sockets.
For "Ci analysis input" optional boards (inductive conductivity),
continue at Step 10; otherwise, at Step 12
10
Connect the inductive conductivity sensor using the M12 socket
adapter from the scope of delivery of the Ci optional board to the
Ci optional board of the device.
11
Connect the 2-wire temperature sensor cable of the M12 socket
adapter to an appropriate analog input (e. g. temperature measur-
ing input). Observe the specifications for the temperature sensor
integrated into the conductivity sensor.
15
Now, reconnect the power supply and check whether the new
hardware has been recognized.
v Chapter 7.2.1 "Checking the optional boards", Page 56
16
Only for Ci optional boards (inductive conductivity):
Perform a Ci base calibration.
v Chapter "Performing the Ci base calibration", Page 11
17
Only for optional boards for analysis inputs:
Calibrate the analysis inputs.
v Chapter 11 "Calibration in general", Page 15