Calibrating Universal Inputs; Important Information; General Information; Calibration Methods For Universal Inputs - JUMO AQUIS touch P Installation Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch P:
Table of Contents

Advertisement

17.1

Important information

17.2

General information

17.2.1

Calibration methods for universal inputs

Universal input
Operation modes
Calibration
routine
Zero-point calibration
Slope calibration
Two-point calibration
Rel. cell constant
Temperature coefficient
TC-curve
Zero-point calibration
Slope calibration
WARNING!
During the calibration, the relays and analog output signals assume the config-
ured states!
Universal inputs can be configured with various operation modes for a number
of different process variables (see table below).
Detailed information on the possible configurations:
v Chapter 9.5.2 "Universal inputs of base unit and optional boards", Page 99
Appropriate calibration routines for each operation mode of a universal input can
be enabled in the calibration default settings.
The following table lists the availability of calibration routines for the individual
operation modes.
linear
scaling
X
X
X
This calibration method is used to determine the pH zero point on a measure-
ment characteristic curve. The slope is retained.
A test solution with a defined value of the respective process variable is needed
as a reference.
This calibration method is used to determine the slope of a measurement char-
acteristic curve. The zero point is retained.
A test solution with a defined value of the respective process variable is needed
as a reference.

17 Calibrating universal inputs

Temp. mea-
pH value
surement
measure-
ment
Conductivity
measure-
ment
X
X
X
X
X
free chlorine,
ph/T-com-
pensated
X
53

Advertisement

Table of Contents
loading

This manual is also suitable for:

202580

Table of Contents