JUMO AQUIS touch P Installation Instructions Manual page 54

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch P:
Table of Contents

Advertisement

17 Calibrating universal inputs
Two-point calibration
Calibrating rel. cell constants, temperature coefficients, and TC-curves
54
The zero point and slope of the measuring characteristic curve are determined
with the aid of 2 measurings of 2 different reference solutions.
Two test solutions with defined values of the respective process variable are
needed as a references.
The calibration routines for conductivity measuring with the universal input cor-
respond to those for the Ci calibration.
v Chapter 16 "Calibrating Ci conductivity sensors", Page 43