Slope Calibration (Free Chlorine, Ph/Temp.-Compensated) - JUMO AQUIS touch P Installation Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch P:
Table of Contents

Advertisement

17.3.3

Slope calibration (free chlorine, ph/Temp.-compensated)

Step
Action
1
Start the slope calibration.
Device Menu r Calibration r Select Universal Input r Slope Cal-
ibration
2
Clean the sensor and immerse it in the test solution.
3
Check the displayed values for the pH value and
temperature. Automatic sensing can be configured in the calibra-
tion default settings for both process variables independently of
one another. With automatic sensing, the particular influencing
variable is only displayed and can no longer be changed here.
Without automatic sensing, the particular influencing variable must
be entered manually here.
Wait until the measuring value displayed stabilizes and then check
the pH value displayed. Then confirm by pressing "OK".
Example with temperature sensing and without pH-value sensing
Measuring value
based on previous
calibration values
Tap button to enter pH
value of test solution
4
Enter the concentration value of the test solution. Confirm by
pressing "OK"
Tap button to enter
chlorine concentration
5
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
17 Calibrating universal inputs
61

Advertisement

Table of Contents
loading

This manual is also suitable for:

202580

Table of Contents