Calibrating A Ph Measuring Chain; Notes; General Information; Calibration Methods For Ph Sensors - JUMO AQUIS touch P Installation Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch P:
Table of Contents

Advertisement

12.1

Notes

12.2

General information

12.2.1

Calibration methods for pH sensors

Zero-point calibration
Two-point Calibration
Three-point calibration

12 Calibrating a pH measuring chain

WARNING!
During the calibration, the relays and analog output signals assume the config-
ured states!
The calibration of pH electrodes is based on measurements in Buffer solutions
with a defined pH-value. The pH values of the Buffer solutions used are speci-
fied either via entry of fixed values into the calibration default settings, entered
during the calibration or recognized automatically by "automatic buffer recogni-
tion" during the calibration process. For "automatic buffer recognition", a Buffer
set table must be selected in the calibration default settings. In this case, the
Buffer solutions used must be listed in the provided Buffer set table. Since pH
value measurement of liquids is temperature-dependent, the temperature of the
Buffer solution must be sensed in order to compensate for its effect on the result
of the measurement. This requires either manual entry or measurement with the
aid of a Temperature sensor.
This calibration method is used to determine the pH zero point on the measure-
ment characteristic curve. The Slope is retained.
A Buffer solution with a defined pH-value is needed as a reference.
By measuring 2 different Buffer solutions with defined pH-values, the pH zero
point and pH slope of the measuring chain are established.
The pH-values of the Buffer solutions must be at least 2 pH apart. This calibra-
tion method is recommended for most applications.
With Three-point calibration, the pH zero point as well as the pH slope in the
acidic region and the pH slope in the alkaline region are established.
This method requires 3 Buffer solutions with defined pH-values as references.
One of these must be acidic, one neutral, and one alkaline. The pH values of the
Buffer solution must be at least 2 pH apart from one another. This calibration
method is recommended for applications with more demanding accuracy re-
quirements when performing measurements in both the alkaline and acidic re-
gions.
21

Advertisement

Table of Contents
loading

This manual is also suitable for:

202580

Table of Contents