18 Technical data
18.12 Case
Site altitude
Case type
Materials
Dimensions of the front
Dimensions of the panel cut-out
Depth behind panel
without Ci-conductivity sensor
with Ci-conductivity sensor
Panel thickness
Minimum distance to other operating
equipment
Ambient temperature
Storage temperature
Resistance to climatic conditions
Operating position
Protection type
Case front in panel insertion
Metal case barrel
Weight
(fully fitted)
76
maximum 2000 m above sea level
Plastic front frame with metal case barrel (for indoor use only)
Plastic front panel UL 94 V0
Case barrel from zinc-plated steel sheet
96 mm × 96 mm
92 mm × 92 mm
Tolerance = +0.8 mm
130.9 mm
283.3 mm (including manoeuvring space for sensor cable)
Max. 5 mm
from edge of the panel cut-out
horizontal 35 mm, vertical 80 mm
-5 to +50 °C
-30 to +70 °C
Relative humidity < 85% annual average, no condensation
Any (with due consideration for the viewing angle of the
screen)
According to DIN EN 60529
IP66
IP20
approx. 1000 g