JUMO AQUIS touch P 202580 Operating Manual
JUMO AQUIS touch P 202580 Operating Manual

JUMO AQUIS touch P 202580 Operating Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Table of Contents

Advertisement

Quick Links

JUMO AQUIS touch P
Modular Multichannel Measuring Device for
Liquid Analysis with Integrated Controller and
Paperless Recorder
Type 202580
Operating Manual
b20.2580.0en
V2.00/EN/00607974

Advertisement

Table of Contents
loading

Summary of Contents for JUMO AQUIS touch P 202580

  • Page 1 JUMO AQUIS touch P Modular Multichannel Measuring Device for Liquid Analysis with Integrated Controller and Paperless Recorder Type 202580 Operating Manual b20.2580.0en V2.00/EN/00607974...
  • Page 2 Caution! If the device or a sensor connected to it fails abruptly, it is likely that a dangerous overdosage has occurred! For this case, appropriate precautionary measures must be taken. NOTE! Read this operating manual before putting the device into service. Keep the operating manual at a location that is readily accessible to all users.
  • Page 3: Table Of Contents

    Content Safety information ............13 Warning symbols ................13 Indicative symbols ................13 Intended use ..................14 Qualification of personnel ..............14 Acceptance of goods, storage, and transport ....15 Checking the delivery ..............15 Important information about storage and transport ......15 Returning goods ................
  • Page 4 Content Connection diagram ................. 40 6.4.1 Overview of connections ..............40 6.4.2 Analog inputs base unit ..............42 6.4.3 Analog inputs optional boards ............44 6.4.4 Analog outputs ................. 52 6.4.5 Binary inputs ..................53 6.4.6 Binary outputs, power supply unit board .......... 55 6.4.7 Binary outputs, optional boards ............
  • Page 5 Content 8.3.2 Event list ..................88 Memory Manager (USB flash drive) ..........90 Operating the controllers ..............93 8.5.1 Automatic control mode ..............94 8.5.2 Controller in the manual mode ............96 8.5.3 Hold mode ..................97 8.5.4 Optimizing controllers ..............98 Operation of the data monitor/recording function ......
  • Page 6 Content 10.5.5 CR/Ci measuring ranges ............... 129 10.6 Analog outputs of base unit and optional boards ......132 10.7 Digital inputs of base unit and optional boards ......133 10.8 Digital outputs of base unit and optional boards ......134 10.9 Limit monitoring and alarms ............
  • Page 7 Content Calibration in general ...........169 12.1 Notes ..................... 169 12.2 General information ............... 169 12.2.1 General procedure for calibration ..........169 12.3 Calibration logbook ................ 171 Calibrating a pH measuring chain .......175 13.1 Notes ..................... 175 13.2 General information ............... 175 13.2.1 Calibration methods for pH sensors ..........
  • Page 8 18.3.3 Slope calibration (free chlorine, ph/Temp.-compensated) ..... 215 PC Setup Program ............217 19.1 General Information ............... 217 19.2 Installing the JUMO PC setup program ......... 218 19.2.1 Procedure ..................218 19.3 Setup program login ..............220 19.3.1 Rights in the setup program ............221 19.4...
  • Page 9 19.7.3 Configuring the connection without the assistant ......248 19.7.4 Searching for a device in the network ..........250 19.8 Setting up the device with the JUMO PC setup program ....252 19.8.1 Configuring and parameterizing ............. 252 19.8.2 User list ..................252 19.8.3 Country settings ................
  • Page 10 Content 19.9.2 Screenshot ..................295 19.9.3 Ethernet ..................295 19.9.4 Enabling of extra codes ..............296 19.9.5 Deleting internal measurement data ..........296 19.9.6 Adjusting / Testing ................. 297 Technical data ...............303 20.1 Analog inputs base unit ..............303 20.1.1 Temperature measuring input (IN 4) ..........303 20.1.2 Temperature measuring input (IN 5) ..........
  • Page 11 Content 20.12 Case ....................316 20.13 Functions ..................317 20.13.1 Controller channels ................ 317 20.13.2 Recording function ................ 318 20.13.3 Customer-specific linearization ............318 20.14 Approvals/approval marks ............. 318...
  • Page 12 Content...
  • Page 13: Safety Information

    1 Safety information Warning symbols DANGER! This symbol indicates that the risk of personal injury from electrocution exists if the appropriate precautionary measures are not taken. WARNING! In conjunction with the signal word "Warning", this symbol indicates that the risk of personal injury or death exists if the appropriate precautionary measures are not taken.
  • Page 14: Intended Use

    1 Safety information Intended use The JUMO AQUIS touch P is is designed for measurement, control, and auto- mation tasks in industrial environments as specified in the technical data. Use for any other purpose is considered contrary to the intended use.
  • Page 15: Acceptance Of Goods, Storage, And Transport

    Description of the application • Description of the error that has occurred The accompanying letter for repair is linked to www.jumo.de on the Internet un- der the heading Service & Support as follows: Product Service r Repair Service r Returning a unit...
  • Page 16: Protection Against Electrostatic Discharge (Esd)

    2 Acceptance of goods, storage, and transport 2.3.2 Protection against electrostatic discharge (ESD) CAUTION! Electrostatic charges occur in non-ESD-protected environments. Electrostatic discharge can damage assemblies or components. For transport purposes, use only the ESD packaging provided. To prevent damage from ESD, electronic assemblies, or components with a high internal resistance must be handled, packaged, and stored in an environ- ment that protects against ESD.
  • Page 17: Device Description

    6 binary signals are registered and displayed on the screen in a time-depen- dent curve. Storage is tamper-proof and enables official recording obligations to be fulfilled. The data can be extracted using the JUMO PCC software or to a USB flash drive for evaluation by the JUMO PCA 3000 PC evaluation software.
  • Page 18 3 Device Description Application examples The modular setup and open structure of the device permits a host of potential applications: • Municipal and industrial water treatment in wastewater treatment plants • Process systems • Drinking and bathing water monitoring • Pharmaceutical water •...
  • Page 19: Block Diagram

    ISFET), redox-, or NH sensors/ electrolytic conductivity (conductive) Conductivity sensors in 2-/4 conductor technology/ electrolytic conductivity (inductive) for inductive Power supply unit and base unit conductivity sensors from JUMO with series connections Expansion slots for AddOn with Optional boards...
  • Page 20: Device Setup

    3 Device Description Device setup Mounting brackets Metal case barrel Rear panel with connection terminals (base unit and optional boards) Housing front LED (voltage supply switched on) Touch screen TFT color screen USB interfaces...
  • Page 21: Identifying The Device Version

    4 Identifying the device version Nameplate The nameplate on the device enclosure identifies the device version. Example of a nameplate Datamatrix code Information symbol (Read documentation! v Chapter 1.1 "Warning symbols", Page 13) Part number Serial number Approval mark Voltage supply Type key It is helpful to become familiar with the technical features of the device prior to commissioning.
  • Page 22: Order Details

    In addition to the standard languages of German, English and, French, 13 ad- ditional languages (e.g. Russian, Chinese, Italian etc.) are available. Please use the contact information on the back of this manual to contact JUMO in this regard. Slot...
  • Page 23 4 Identifying the device version (7) Input/output 2 IN 12, OUT 8/9 Not used Universal input Relay (changeover contact) 2× relays (normally open contact) Solid state relay triac 230 V, 1 A Logic output 0/22 V 2× logic outputs 0/12 V Analog output 2×...
  • Page 24 Panel seal Mini-DVD with JUMO PC setup program as demo version, Adobe Acrobat Reader, operating man- ual and data sheet in PDF format, GSD generator and JUMO PCC / PCA3000 as demo version Installation instructions in 2 volumes B 202580.4...
  • Page 25 00592963 USB flash drive 2.0 (1 GB) 00505592 USB cable, A-connector to Mini-B connector, 00506252 length 3 m JUMO PC setup program AQUIS touch S/P, 00594355 (PG202599) JUMO PCA3000/PCC software package 00431884 PhotoMOS® is a registered trademark of Panasonic. The USB flash drive indicated has been tested and is designed for industrial applications. No liability is assumed for flash drives from other manufacturers.
  • Page 26 4 Identifying the device version...
  • Page 27: Mounting

    5 Mounting Notes DANGER! Under no circumstances may the device be installed or removed while under voltage! This poses the risk of electrocution. Switch-off the entire system beforehand. This work must be performed only by qualified personnel! The device must never be installed in potentially-explosive areas! There is the risk of an explosion.
  • Page 28: Dimensions

    5 Mounting Dimensions 130,9 mm 123,1 mm 96 mm Panel thickness max. 5 mm 22,25 mm 22,25 mm 5,5 mm +0.8 mm 92 mm Spacings For adequate ventilation, the following mounting distances from adjacent equip- ment must be respected: • horizontal 35 mm •...
  • Page 29 5 Mounting Mounting depth with Ci-conductivity sensor 283,3 mm...
  • Page 30: Panel Insertion

    The insertion operation is described below. NOTE! Insertion in a control panel provides protection class IP66. When the JUMO AQUIS touch P is inserted in the wall of control cabinets with a protection class higher than IP66, the protection class of the control cabinet lapses and corre- sponds to the protection class for control panel insertion (IP66).
  • Page 31 5 Mounting Metal case barrel of the JUMO AQUIS touch P Mounting elements Seal from the scope of delivery for the device Case front Panel/control cabinet Procedure Step Action Push the panel seal provided (3) from the rear panel of the device over the case barrel to the front panel.
  • Page 32 5 Mounting...
  • Page 33: Electrical Connection

    6 Electrical connection Installation notes DANGER! Observe the following instructions! Qualification of personnel • The electrical connection must only be carried out by qualified personnel. Electrical wiring • When selecting the electrical wiring material as well as when installing and connecting the device electrically, comply with the requirements of DIN VDE 0100 "Low-voltage electrical installations"...
  • Page 34 6 Electrical connection • The electromagnetic compatibility conforms to the standards and regulations cited in the technical data. • The USB interfaces and digital inputs are not electrically isolated. The information regarding electrical isolation must be observed. v Chapter 6.3 "Galvanic isolation", Page 39...
  • Page 35: Connecting Cables

    6 Electrical connection Connecting cables 6.2.1 Preparing coaxial cables for pH/redox electrodes ® Coaxial cable with Shield-Kon connector Length Part number 1.5 m 00085154 00307298 10 m 00082649 Preparing your own coaxial cable 25 mm Sensor Remove outer sheathing from the cable r Pull back the braided shield Inner conductor Shield-Kon®...
  • Page 36 6 Electrical connection Inner conductor Shrink sleeve Use shrink tubing to insulate the braided shield NOTE! The black, semiconducting layer must not touch with the inside conductor! This would short-circuit the signal from the pH electrode. Preparing a shielded multiconductor cable 60 mm Strip the insulation off the connecting cable and pull back the shield 60 mm...
  • Page 37: Conductor Cross-Sections For Base Unit And Power Supply Unit

    6 Electrical connection 60 mm Shield-Kon® connector Use shrink tubing to insulate the shield and Shield-Kon® connector 6.2.2 Conductor cross-sections for base unit and power supply unit The terminals on the base unit and power supply unit are plug-in screw termi- nals.
  • Page 38: Conductor Cross-Sections For Optional Boards

    6 Electrical connection 6.2.3 Conductor cross-sections for optional boards The terminals on optional boards are plug-in screw terminals. Optional boards for Ferrule Conductor cross sec- Length to tion strip minimal maximal Universal inputs without ferrule 0.14 mm 1.5 mm 7 mm Analog outputs with ferrule with lip 0.25 mm...
  • Page 39: Galvanic Isolation

    If sensors that are not electrically isolated are connected to a digital input and supplied by an external power source, potential differences between the in- ternal and external ground can cause problems. Providing the supply voltage from the voltage supply outputs of the JUMO AQUIS touch P is preferable in such cases.
  • Page 40: Connection Diagram

    6 Electrical connection Connection diagram 6.4.1 Overview of connections Options Base unit Supply unit Module Connector terminal Type Inputs Base IN 1 to IN 3 Binary inputs unit IN 4 to IN 5 Temperature inputs IN 6 Universal input Option- IN 7 to IN 8 analysis inputs boards...
  • Page 41 6 Electrical connection Outputs Power OUT 1 to 2 Digital outputs (relay changeover contact) supply unit Base OUT 3 Not available unit OUT 4 Analog output OUT 5 Not available – Option- OUT 6 to OUT 13 Analog/digital outputs, OUT 8/9 also for voltage supply output DC ±5 V, 24 V boards Interfaces...
  • Page 42: Analog Inputs Base Unit

    6 Electrical connection 6.4.2 Analog inputs base unit Connector/ Connection variant Symbol terminal IN 4 RTD temperature probe 2-wire circuit Pt100, Pt1000 or customer-specific characteristic line RTD temperature probe 3-wire circuit Pt100, Pt1000 or customer-specific characteristic line IN 5 RTD temperature probe 2-wire circuit Pt100, Pt1000 or customer-specific characteristic line...
  • Page 43 6 Electrical connection Connector/ Connection variant Symbol terminal IN 6 Standard signal Current 0(4) to 20 mA...
  • Page 44: Analog Inputs Optional Boards

    6 Electrical connection 6.4.3 Analog inputs optional boards Universal inputs Slot Connection variant Symbol IN 11 RTD temperature probe IN 12 2-wire circuit Pt100, Pt1000 or customer-specific characteristic line RTD temperature probe 3-wire circuit Pt100, Pt1000 or customer-specific characteristic line Resistance potentiometer A = Start E = End...
  • Page 45 The conductor colors listed refer to JUMO ISFET-pH sensors. The orange-colored conductor is not con- nected. The optional board "Voltage supply output DC ±5 V, 24 V“ (part no. 592963) is required for the voltage supply to the JUMO ISFET-pH sensor.
  • Page 46 6 Electrical connection When connecting the temperature probe of JUMO ISFET pH sensors to process connection 615 (NTC 8k55), no customer-specific linearization like that for the JUMO AQUIS 500 pH is needed. Temperature input IN 5 supports connection of 8k55-NTC temperature probes.
  • Page 47 6 Electrical connection Slot Option/connection variant Symbol IN 7 pH/Redox IN 8 Asymmetric connection of a combination electrode with integrated RTD temperature probe and VarioPin terminal head The RTD temperature probe is used to provide a tempera- ture-compensated pH-value measurement, and can be con- nected to a temperature input or universal input.
  • Page 48 6 Electrical connection Slot Option/connection variant Symbol IN 7 pH/Redox IN 8 Symmetric connection of a combination electrode Symmetric connection is used to reduce interference from stray electromagnetic fields along the sensor cable. A = Glass/metal electrode (core) B = Reference electrode (inner shield) C = Liquid potential (grounding pin, pipe, or container wall at the measuring point) D = Shield (outer shield)
  • Page 49 6 Electrical connection Slot Option/connection variant Symbol IN 7 pH/Redox IN 8 Symmetric connection of a combination electrode with integrated resistance thermometer and VarioPin plug head Symmetric connection is used to reduce interference from stray electromagnetic fields along the sensor cable. The RTD temperature probe is used to provide a tempera- ture-compensated pH-value measurement, and can be con- nected to a temperature input or universal input.
  • Page 50 2-electrode system with 2-wire conductor Terminal 1 must be connected to the outer electrode on concentric conductivity sensors. A = Outer electrode (core color for JUMO types with fixed ca- ble: White) B = Inner electrode (core color for JUMO types with fixed ca-...
  • Page 51 IN 8 4-electrode system A = Outer electrode 1 (I hi) (core color of CR-4P cable for JUMO types: Red) B = Inner electrode 1 (U hi) (core color of CR-4P cable for JUMO types: Gray) C = Inner electrode 2 (U lo) (core color of CR-4P cable for...
  • Page 52: Analog Outputs

    6 Electrical connection 6.4.4 Analog outputs Base unit Connector/ Connection variant Symbol terminal OUT 4 Analog output DC 0 to 10 V or DC 0(4) to 20 mA (configurable) Optional boards Slot Option/connection variant Symbol OUT 6/7 Analog output OUT 8/9 DC 0 to 10 V or OUT 10/11 DC 0(4) to 20 mA...
  • Page 53: Binary Inputs

    6 Electrical connection 6.4.5 Binary inputs Base unit Connec- Connection variant Wire Potential Terminal Symbol tor/ DC 24 V IN 1 IN 2 IN 3 terminal OUT 8/9 IN 1 to 3 Digital input Potential-free contact (potential-free contact) In the digital input configuration, the "Contact" option must be set to "Potential-free contact".
  • Page 54 6 Electrical connection Connec- Connection variant Wire Potential Terminal Symbol tor/ DC 24 V IN 1 IN 2 IN 3 terminal OUT 8/9 IN 1 to 3 Digital input Sensor + (NPN transistor Sensor - switching output) Switching signal (collector) Sensor - In the digital input configuration, the "Contact"...
  • Page 55: Binary Outputs, Power Supply Unit Board

    6 Electrical connection Optional boards Connector/ Connection variant Symbol terminal IN 13/14/15 3× binary input IN 16/17/18 6.4.6 Binary outputs, power supply unit board Connector/ Connection variant Symbol terminal OUT 1 Relay OUT 2 Changeover contact...
  • Page 56: Binary Outputs, Optional Boards

    6 Electrical connection 6.4.7 Binary outputs, optional boards Slot Option/connection variant Symbol OUT 6/7 Relay OUT 8/9 Changeover contact OUT 10/11 OUT 12/13 2× relay Normally open contact Solid state relay triac 230 V/1 A 2× solid state relay PhotoMOS® 50 V/200 mA Binary output 0/22 V...
  • Page 57: Mains Power Connection

    6 Electrical connection WARNING! Combining a mains voltage circuit with a protective low-voltage circuit on the option "dual normally open contacts" is not permissible. 6.4.8 Mains power connection Connector/ Connection variant Symbol terminal PWR IN Mains power input...
  • Page 58: Voltage Supply Outputs

    6 Electrical connection 6.4.9 Voltage supply outputs Optional board Slot Connection variant Symbol Voltage supply DC ±5 V OUT 8/9 for ISFET sensors Voltage supply DC 24 V for external transmitters 24 V...
  • Page 59: Interfaces

    6 Electrical connection 6.4.10 Interfaces Base unit interfaces Connector/ Connection variant Symbol terminal COM 1 RS485 RxD/TxD+ RxD/TxD- Connector/ Connection variant Symbol terminal USB device USB device Type Mini-B (socket) USB host USB host Type A (socket)
  • Page 60 6 Electrical connection Optional board interfaces Slot Connection variant Terminating resistors Symbol COM 2 RS422 with terminating resistors RxD+ terminating resistors with RxD- DIP switches on TxD+ optional board configurable without terminating resistors TxD- RS485 RxD/TxD+ terminating resistors with RxD/TxD- DIP switches on optional board configurable PROFIBUS-DP...
  • Page 61: Startup

    When the battery is empty a battery alarm will indicate that status. The battery must be exchanged before it is empty. The bat- tery must be exchanged by the JUMO Service department! In this case, send in the device!
  • Page 62: Function Test

    In such cases, check whether the optional board con- cerned was installed correctly. v Chapter 11 "Retrofitting optional boards", Page 161 If this is not successful, contact Technical Support at JUMO. The contact data can be found on the back of these installation instructions. 7.2.2...
  • Page 63 The device calculates the (compensated) values of the process variable from the (uncompensated) raw measurement data. Sample screen: IN 7 measures conductivity IN 8 measures pH value IN 9 to 10 are available only on the JUMO AQUIS touch S.
  • Page 64 7 Startup...
  • Page 65: Operation

    The passwords in the device can be changed. v Chapter 8.2.1 "Log-on/Log-out", Page 78 The JUMO PC setup program is needed to change user names and user rights. v "User list", page 252 The following table provides an overview of the factory-set user accounts.
  • Page 66 8 Operation Factory-set user rights Users Every- User 2 User 1 Service Master User rights Display of: • Current measuring values on overview and detailed screens • Configuration data • parameters • Device information Display of: • History of measurement data in the recording function •...
  • Page 67: Display And Control Elements

    8 Operation 8.1.2 Display and control elements (4)(5) (7) (8) Touchscreen Toolbar with buttons for operation "Device menu" button with display of: • Date and time • Logged-in user ("Master" in the example) • Remaining memory display in percent for recording function (in the example: 100%) "Alarm/Event List"...
  • Page 68: Menu Structure

    8 Operation 8.1.3 Menu structure At the operating level, 3 different navigation buttons are available to select ap- propriate screens for display and control of device functions. The "Device menu" and "Alarm/event list" menu levels are also displayed by means of corresponding buttons. The device menu contains submenus for set- ting, servicing, and diagnosing the device and its functions.
  • Page 69 8 Operation Overview of the menu structure Operating level Home Select Next operator operator display display Navigation in the operating loop Plug-in USB flash drive Device settings Memory Alarm-/ menue manager event list Secure hardware Login/logoff Alarm list remove Schreiber Calibration Event list Update auf USB...
  • Page 70 8 Operation Operation screens in the operation loop (16) (15) (14) (13) (12) (11) (10) Fig. Operation screen Description General screen 1 Freely configurable overview display of measuring values and digital sig- nal states General screen 2 The overviews can be configured as a 2-part screen or a 4-part screen. 2-part screen: Display of 2 primary and 2 secondary measured values per screen, 1 ad- ditional measured value and 3 binary values.
  • Page 71 8 Operation Fig. Operation screen Description Detailed screen 1 Freely configurable large-section screen Detailed screen 2 Display of 1 main measuring value, 1 second measuring value, 1 addi- Detailed screen 3 tional value and up to 3 binary values, as well as additional visualization Detailed screen 4 of the main measuring value and the alarm limits for an analog input with Detailed screen 5...
  • Page 72 8 Operation Display of invalid values Invalid input signals/measured values or errors in the analog input configuration are detected and shown in the measured value displays as follows: Type of error Display underrange: measuring range underflow overrange: measuring range overflow Compensation error: An error occurred when compensat- ing for factors affecting the measurements of the analysis.
  • Page 73: Entry Of Text And Numbers

    8 Operation 8.1.4 Entry of text and numbers Dialogs for entry of text or numbers appear automatically when the correspond- ing entry field is tapped. Text entry dialog There are 2 special features in addition to conventional entry of characters: •...
  • Page 74 8 Operation Number entry dialog This dialog opens if an entry field for numerical values is tapped. Special feature : The "Exp" button permits entry of an exponent as a power of ten. Procedure: Enter the numerical base value r Tap "Exp" r Enter the exponent r Confirm the entry Entry dialog buttons Explanation...
  • Page 75: Device Menu

    8 Operation Device menu NOTE! Operation depends on the user's rights. Operating and setting options are re- stricted, depending on the user who is logged in. The "Master" and "Service" users have access to all menus and functions (fac- tory setting). v Chapter 8.1.1 "Passwords and user rights", Page 65 The device menu contains submenus for setting and configuring all device func- tions.
  • Page 76 No user level is configured at the factory. The user level must be configured and loaded into the device with the aid of the JUMO PC setup program. The "User Level" entry does not appear in the device menu as long as no user level is config- ured.
  • Page 77 8 Operation To navigate in submenus, menu items are opened by tapping the folder icons identified with a plus sign. Open menu structures are identified with a minus sign and can be closed again by tapping the folder icon. Currently open windows can be left either by tapping "Exit" or the "Close win- dow"...
  • Page 78: Log-On/Log-Out

    The log-on/log-out dialog is open • The memory manager is open • During calibration of analysis sensors • During calibration of the touchscreen The re-authentication time is set using the JUMO PC setup program. v Chapter 19.8.2 "User list", Page 252...
  • Page 79: User Level

    User level The user level consists of a user-defined list of parameters and configuration settings. You need the JUMO PC setup program to configure the user level. v Chapter 19.8.5 "User level", Page 263 The user level can be opened from the device menu only if it has been config- ured beforehand with the aid of the PC setup program.
  • Page 80: Functional Level

    8 Operation 8.2.3 Functional level The functional level is used primarily for testing and diagnostic purposes. Analog and binary values of the outputs can be controlled manually here. This may be useful, for instance, for checking an individual piece of equipment in a plant. When performing maintenance and repair work, counters for operating hours, switching actions and flow rates can be reset.
  • Page 81 8 Operation Manual control of analog/digital outputs Activate the configuration setting "Enable menu mode" for the outputs you wish to control manually. "Manual mode buttons" buttons for manual control for these outputs then appear at the functional level. To adjust output values manually, proceed as follows: Select Device menu r Functional Level r Analog or Digital Output tab r Tap "Manual mode: button r Tap Output Display field r Entry dialog...
  • Page 82: Device Information

    8 Operation 8.2.4 Device information For testing and diagnostic purposes, the "Device info" menu provides access to extensive data about the hardware and software of the device, as well as current analog and binary values. Device Info menu items: • General: Information about the main board, device software and Ethernet configuration •...
  • Page 83: Service

    Chapter 8.4 "Memory Manager (USB flash drive)", Page 90 • Debug Window: The Debug window can be used to check the behavior of the device software in detail. Data are displayed to assist the JUMO service technicians during troubleshooting. •...
  • Page 84: Calibrating The Touchscreen

    When the battery is empty a battery alarm will indicate that status. The battery must be exchanged before it is empty. The bat- tery must be exchanged by the JUMO Service department! In this case, send in the device! 8.2.6...
  • Page 85: Alarm/Event List

    8 Operation Alarm/Event list The JUMO AQUIS touch P offers the option of configuring alarm functions and event functions in numerous functions. In addition, the electronics of the JUMO AQUIS touch P are self-monitoring and trigger corresponding preprogrammed alarms and events in case the internal device malfunctions.
  • Page 86: Alarm List

    8 Operation 8.3.1 Alarm list The alarm list displays the current alarms. Alarms are cleared upon elimination of the alarm condition. Each alarm triggers a "collective alarm". The alarm list view contains buttons for viewing details about alarms and acknowledging col- lective and dosing alarms.
  • Page 87 8 Operation Acknowledging collective/dosing alarms The collective alarm combines all alarms in the alarm list. It simplifies signaling of one or several active alarms with external indicating devices or to control rooms. The digital signal for the collective alarm is available in the binary selec- tor in 2 versions: •...
  • Page 88: Event List

    8 Operation 8.3.2 Event list A number of situations that are essential for tracking and diagnostic purposes are logged in the event list. The entries are identified with corresponding icons on the basis of the type of event. In addition, events are logged with an icon in the data monitor/recording function.
  • Page 89 8 Operation The following table provides an overview of the possible entries in the event list. Events Symbol Power on Power off Alarm occurred Alarm cleared • Configured event (condition occurred) • Calibration start • Timer start • Wash contact start •...
  • Page 90: Memory Manager (Usb Flash Drive)

    The measurement data are stored in DAT files and the configuration data in SET files. This data can be opened and evaluated with the aid of the JUMO PCA3000 evaluation software. Data that has been read out is marked inter- nally as retrieved and the available memory display is reset to 100%.
  • Page 91 8 Operation PCA3000 evaluation software. In contrast to Recorder Update, there is no internal marking of recorder data and no resetting of the available memory display. NOTE! The "Recorder Update" and "Recorder Backup" functions are available only if the extra code "Recording" is enabled. NOTE! A measuring data recording session is closed by changing configuration data that are relevant to the data monitor or registration function (e.g.
  • Page 92 Software Update: The device software can be updated with the aid of a USB flash drive. To do so, an appropriate update file must have been saved the flash drive in advance; this file can be obtained from JUMO Service. CAUTION! It is strongly recommended that the configuration and recorder data be backed prior to performing a software update.
  • Page 93: Operating The Controllers

    8 Operation Operating the controllers NOTE! Since the automatic control mode is of prime importance in the case of control- lers, correct configuration of the individual controller and its parameterization (ad- justment of the control response) are very important for achieving good pro- cess value stability.
  • Page 94: Automatic Control Mode

    Setpoint 1 is active by default. Setpoint 2 can be activated instead of Setpoint 1 in each controller by means of "Setpoint selection". Setpoints can also be spec- ified by external sources and transmitted to the controllers in the JUMO AQUIS touch S via analog inputs. "Setpoint selection" and the configuration of external setpoints are set in the "Setpoint configuration".
  • Page 95 8 Operation On the operation screen of each controller, it is possible to change the current controller setpoint value, switch to the "Manual mode" or start "Self-optimiza- tion". Display of the digital controller output signals as indicator lights "Manual entry" button for changing the current setpoint value A change to the currently active setpoint value is transferred to the con- troller parameters.
  • Page 96: Controller In The Manual Mode

    8 Operation 8.5.2 Controller in the manual mode In manual mode, the user can manually control the controller outputs. The controller outputs can be controlled in two ways: • Entry of a numerical value: Using the "Manual entry" button, the user can open a dialog to enter a fixed numerical value for the output. •...
  • Page 97: Hold Mode

    8 Operation 8.5.3 Hold mode The hold mode of a controller is activated in two ways: • Calibration of the actual-value input • Hold signal for the individual controller (specified in the controller configura- tion) In the hold mode, automatic control is suspended. The controller generates the preconfigured value as output level unless acceptance of the hold output is dis- abled in the configuration.
  • Page 98: Optimizing Controllers

    8 Operation 8.5.4 Optimizing controllers The control response can be optimized by entering known controller parameters manually or automatically by means of "self-optimization". During self-optimiza- tion, the individual controller determines the mathematical parameters for a pro- cess. The controller changes the output level (step change) and evaluates the response of the actual value in the process (step response).
  • Page 99: Operation Of The Data Monitor/Recording Function

    8 Operation Operation of the data monitor/recording function The standard version of the JUMO AQUIS touch P is equipped with a data mon- itor. It is used to record and view the analog measuring value and signal states of digital functions. Two groups are available, each of which can record up to 4 analog values and 3 binary values and display them in the form of a line recorder diagram.
  • Page 101 8 Operation Stylus trace view The "Next channel" button is used to scroll through the individual stylus trace views for the channels. For the particular channel selected, the stylus trace view shows the corresponding scale with a stylus. The position of the stylus corresponds to the current measuring value of the channel.
  • Page 102 8 Operation Scrolling through the stylus trace views The sequence of views that appear with repeated tapping of the "Next Channel" button can be seen in the graphic below. General line recorder screen Stylus trace view Stylus trace view Channel 4 Channel 1 Channel advance...
  • Page 103: History Function

    8 Operation 8.6.2 History function The history function is available only in the recording function. It permits the user to view all recorded data in the ring memory on the device's display. The history is opened by tapping the "History" button on the operation screen for the respective group.
  • Page 104: Online Visualization

    Chapter 19.8.13 "Web server", Page 291 The online visualization is open with a web browser. Enter either the IP address or the URL of the JUMO AQUIS touch P in the ad- dress bar of your web browser. v Chapter 10.16 "Ethernet", Page 153 When the website opens, it is possible to select between "Visualization"...
  • Page 105 8 Operation The visualization displays a view identical in appearance to the device. A re- quest for a password appears next. The web server password specified in the web server configuration must be entered here. v Chapter 19.8.13 "Web server", Page 291 As on the device, an operation screen can now be selected from the operation loop.
  • Page 106 The open Quad View displays 4 plus signs for this purpose. Clicking on any one opens a request to enter the IP address of the JUMO AQUIS touch P. After the IP address has been entered, the selected view opens and can be con- trolled exactly as in the "Visualization"...
  • Page 107 8 Operation Example of a Quad View:...
  • Page 108 8 Operation...
  • Page 109: Parameterization

    • Switch time hh:mm:ss Time to switch to standard time NOTE! The date and time are set in the online parameters in the JUMO PC setup pro- gram. v See chapter 19.9.1 "Date and time" Page 294...
  • Page 110: Parameter Sets (Controller Parameter)

    9 Parameterization Parameter Sets (Controller parameter) The parameters for the controller channels determine the control response of the individual control circuits. For stable control response, these parameters must be matched to the prevailing process conditions. Each controller channel has two parameter sets; the selection is made by means of a binary signal. "Pa- rameter set 1"...
  • Page 111 9 Parameterization Parameter Setting op- Explanation (Formula sym.) tions Switching peri- 0 to 9999 s If an output of a controller channel is con- od 1 figured as a Pulse length output, the period factory setting: between the switching pulses is fixed. 20.0 s The length of the switching period should be selected such that, on the one hand, the...
  • Page 112 9 Parameterization Parameter Setting op- Explanation (Formula sym.) tions min. relay 0 to 60 s Depending on the Min. relay switch-on switch-on time 1 time, a lower limit is set for the pulse length factory setting: or an upper limit is set for the pulse fre- 0.0 s quency.
  • Page 113: Setpoint Values

    9 Parameterization Setpoint values The "Setpoint values" submenu allows two setpoint values for each controller channel to be entered. As with the parameter sets, it is also possible to switch between setpoint values by means of a binary signal. "Setpoint value 1" is active by default..
  • Page 114 9 Parameterization...
  • Page 115: Configuration

    NOTE! Changes to the configuration settings described in this chapter can be made directly on the device or via the JUMO PC setup program. NOTE! Settings in the "Configuration" menu can be changed only if a user with corre- sponding user rights is logged in.
  • Page 116: Basic Settings

    0 to 100% If the available memory display reaches this val- ue, the memory alarm is triggered. The following settings can be edited only JUMO PC with the aid of the setup program Setup quick info up to 20 text...
  • Page 117: Display

    10 Configuration 10.3 Display 10.3.1 General Information Open: Device menu r Configuration r Display r General Configuration point Selection/ Explanation setting option Lock touchscreen Selection from Digital signal that blocks operation of the touch- binary selector screen (e.g. key switch for locking operation) Simulation of When this function is activated, alternating on/off inputs...
  • Page 118: Screen

    10 Configuration 10.3.2 Screen Open: Device menu r Configuration r Display r Screen Configuration item Selection/ Explanation setting option Activation of screen- inactive Type of screensaver activation saver time to switch off control signal Wait time for screen- 10 to 32767 s only with activation of the screensaver after saver wait time:...
  • Page 119: Operating Loop

    10 Configuration 10.4 Operating loop 10.4.1 General screens Open: Device menu r Configuration r Operation Loop r General screen r General screen 1 to 2 Configuration point Selection/ Explanation setting option Gener. screen type 2-part screen Selection of the type of overview screen; 4-part screen 2-part screen: Display of 2 main values, 2 sec- ond values, 1 additional value and 3 binary val-...
  • Page 120: Detailed Screens

    10 Configuration 10.4.2 Detailed screens Open: Device menu r Configuration r Operation Loop r Individual screen r Individual screen 1 to 6 Configuration point Selection/ Explanation setting option Screen title up to 31 text characters Title of the individual screen Input signal main val- Selection from Signal source of the analog value displayed as...
  • Page 121: Analog Inputs

    Each value pair assigns a display value (Y-column) to a measuring value (X-column). Up to 8 linearization tables can be stored. To cre- ate a linearization table, you need the JUMO PC setup program. Start of -99999 to +99999...
  • Page 122 10 Configuration Configuration point Selection/ Explanation setting option 0 to 99999 Ω only for IN 5: Resistance value between the slid- er (S) and start point (A) in a resistance transmit- ter/potentiometer when the slider is at the start point. 6 to 99999 Ω...
  • Page 123: Universal Inputs Of Base Unit And Optional Boards

    10 Configuration 10.5.2 Universal inputs of base unit and optional boards Base unit universal input: IN 6 Optional board universal inputs: IN 11/12 Open: Device menu r Configuration r Analog inputs r Universal input 1 to 3 r Configuration Configuration point Selection/ Explanation setting option Description...
  • Page 124 Each value pair assigns a display value (Y-column) to a measuring value (X-column). Up to 8 linearization tables can be stored. To cre- ate a linearization table, you need the JUMO PC setup program. Unit up to 5 text characters...
  • Page 125 10 Configuration Configuration point Selection/ Explanation setting option 6 to 4000 Ω only for IN 11/12: Span of the variable resistance value between the slider (S) and start point (A) 0 to 4000 Ω only for IN 11/12: Resistance value between the slider (S) and end point (E) in a resistance ther- mometer/resistance potentiometer when the slid- er is at the end point.
  • Page 126: Ph/Redox/Nh Analysis Inputs

    10 Configuration 10.5.3 pH/Redox/NH analysis inputs Open: Device menu r Configuration r Analog inputs r Analysis input 1 to 4 r Configuration Configuration point Selection/ Explanation setting option Description up to 15 text characters Designation for the input Electrode type pH standard Type of electrode connected pH antimony...
  • Page 127 10 Configuration Configuration point Selection/ Explanation setting option Text sensor alarm up to 21 text characters Text for the alarm/event list in case of sensor er- Alarms 1/2 Analog input alarms are used to monitor measured values in relation to adjustable limit values. The alarm settings for all analog device functions are explained together.
  • Page 128: Cr/Ci Analysis Inputs (Conductive/Inductive Conductivity)

    10 Configuration 10.5.4 CR/Ci analysis inputs (conductive/inductive conductivity) Open: Device menu r Configuration r Analog inputs r Analysis input 1 to 4 r Configuration Configuration point Selection/ Explanation setting option Description up to 15 text characters Designation for the input Compensation tem- Selection from Analog input of the compensation thermometer...
  • Page 129: Cr/Ci Measuring Ranges

    10 Configuration Configuration point Selection/ Explanation setting option Measuring ranges Four ranges each can be configured for conduc- 1 to 4 tive/inductive (CR/Ci) conductivity measure- ments. The settings for all CR-/Ci analysis inputs are explained together. v "CR/Ci measuring range configuration", page Alarms 1/2 Analog input alarms are used to monitor measured values per measuring range...
  • Page 130 10 Configuration CR/Ci measuring range configuration Open: Device menu r Configuration r Analog inputs r Analysis input 1 to 4 r Measuring range 1 to 4 Configuration point Selection/ Explanation setting option TDS factor 0.01 to 2.00 only for conductive conductivity with TDS compensation: Conversion factor from measured conductivity to display unit (see configuration item "Unit"...
  • Page 131 Each value pair assigns a display value (Y-column) to a measuring value (X-column). Up to 8 linearization tables can be stored. To cre- ate a linearization table, you need the JUMO PC setup program. Start of -99999 to +99999...
  • Page 132: Analog Outputs Of Base Unit And Optional Boards

    10 Configuration 10.6 Analog outputs of base unit and optional boards Open: Device menu r Configuration r Analog outputs r Analog output 1 to 9 Configuration point Selection/ Explanation setting option Description up to 15 text characters Designation for the output Signal Selection from Analog signal source of the output...
  • Page 133: Digital Inputs Of Base Unit And Optional Boards

    10 Configuration Configuration point Selection/ Explanation setting option Response at hold Specification of the analog output value when the high hold function is activated, during calibration of NAMUR low one of the sensors for the particular output or at NAMUR high error (overrange/underrange) hold safety value...
  • Page 134: Digital Outputs Of Base Unit And Optional Boards

    10 Configuration 10.8 Digital outputs of base unit and optional boards Open: Device menu r Configuration r Digital outputs r Digital output 1 to 17 Configuration point Selection/ Explanation setting option Description up to 21 text characters Designation for the output Signal Selection from Digital signal source for the output...
  • Page 135: Limit Monitoring And Alarms

    10 Configuration 10.9 Limit monitoring and alarms 10.9.1 Limit monitoring Open Limit Monitoring: Device menu r Configuration r Limit Monitoring r Limit monitoring 1 to 8 Configuration point Selection/ Explanation setting option Description up to 21 text characters Designation for the limit monitoring Input signal Selection from Signal source of the analog value that is...
  • Page 136 10 Configuration Configuration point Selection/ Explanation setting option Response at hold inactive Specification of the alarm state when the hold active function is activated, at calibration of the particu- hold lar input or at error (overrange/underrange) Response at inactive inactive: alarm suppressed calibration active hold...
  • Page 137 10 Configuration Minimum alarm (On-signal when value drops below lower limit) Binary value Hysteresis Input signal Limit value Maximum alarm(On-signal when value exceeds upper limit) Binary value Hysteresis Input signal Limit value Alarm window (On-signal within a configurable value range Binary value Hysteresis...
  • Page 138: Digital Signal Alarms

    10 Configuration 10.9.3 Digital signal alarms Open: Device menu r Configuration r Digital inputs r Digital input 1 to 9 Configuration point Selection/ Explanation setting option Digital signal for hold Selection from the Digital signal for activating the hold function binary selector When the hold function is activated, the alarm as- sumes the state defined in the "Response at...
  • Page 139: Calibration Timers

    10 Configuration 10.10 Calibration timers Every analysis input and universal input has its own calibration timer. Calibration timers signal when sensor calibration is due. Once a particular input has been calibrated successfully, its calibration timer is reset. The signal can be generat- ed, for instance, by digital outputs, external indicator lights or the alarm/event list.
  • Page 140: Controller

    10 Configuration 10.11 Controller 10.11.1 Configuration of the controllers Open: Device menu r Configuration r Controller r Controller 1 to 4 r Configuration Configuration point Selection/ Explanation setting option Controller type Two-point controller Selection of controller type Three-state controller, Coarse-/fine controller, Modulating controller, continuous controller with integrated...
  • Page 141 10 Configuration Configuration point Selection/ Explanation setting option Man. output level 2 0 to 100% only for coarse/fine controllers: Preconfigured Man. output level for the 2nd con- troller output Accepted automatically upon activation of the manual mode mode if acceptance of the manual output level is set to "Yes";...
  • Page 142 10 Configuration Configuration point Selection/ Explanation setting option Alarm ack. Inactive Activation/deactivation of the acknowledgment Active function Dosing alarms of the respective controller must be acknowledged in the "Alarm list" when "Alarm acknowledgement" is activated. Dosing alarms are not cleared automatically when the control deviation drops to a value that is less than or equal to the alarm tolerance.
  • Page 143 For more detailed information on the subject of control technology, you can download the technical article "Control technology - Basic principles and tips for the practitioner" (FAS 525) as a PDF document for free from the JUMO website.
  • Page 144: Controller Inputs

    10 Configuration 10.11.2 Controller inputs Open: Device menu r Configuration r Controller r Controller 1 to 4 r Input Configuration point Selection/ Explanation setting option Description up to 15 text characters Designation of the controller input Actual value Selection from Selection of the analog signal source for the ac- Analog selector tual value...
  • Page 145: Disturbance Feedforward Control

    10 Configuration 10.11.3 Disturbance feedforward control Open: Device menu r Configuration r Controller r Controller 1 to 4 r Disturbance feedforward control Configuration point Selection/ Explanation setting option additive var. disturb. Selection from Analog input for the additive variable disturbance Analog selector The additive output level component added.
  • Page 146: Self-Optimization

    10 Configuration Configuration point Selection/ Explanation setting option Multiplicative Selection from Analog input for the multiplicative variable distur- Disturbance Analog selector bance The ratio of the variable disturbance value at the working point of the variable disturbance is multi- plied by the proportional gain of the controller. Changes in the variable disturbance affect the total gain of the controller.
  • Page 147: Setpoint Value Configuration

    10 Configuration 10.12 Setpoint value configuration Open: Device menu r Configuration r Setpoint Value Configuration r Setpoint value configuration, controller 1 to 4 Configuration point Selection/ Explanation setting option external Selection from Selection of the analog value as setpoint value Setpoint 1 to 2 Analog selector source...
  • Page 148: Timers

    10 Configuration 10.13 Timers Open timer: Device menu r Configuration r Timer r Timer 1 to 10 r Timer Configuration point Selection/ Explanation setting option Timer function Inactive Operating principle of the timer Timers Control timer Control timer: Functions as weekly time switch Settings for a time program during the week v Chapter 10.13.1 "Control timer", Page 150 Timer: Functions as timing element;...
  • Page 149 10 Configuration Configuration point Selection/ Explanation setting option Signal Selection from Selection of an analog signal the deviation of Tolerance band x Analog selector which from the "Tolerance band w signal" is to be monitored The timer starts and continues running only if the amount of deviation is not greater than the set window range.
  • Page 150: Control Timer

    10 Configuration Timing diagram Timer start Timer time After-run time delay Timer end Timer start Timer time After-run time delay Timer end Signal Timer start Output signal After-run time Timer 10.13.1 Control timer Timer settings for a timed program during the week Prerequisite: The "Timer function"...
  • Page 151: Washtimer

    10 Configuration 10.14 Washtimer Open: Device menu r Configuration r Wash timer r Wash timer 1 to 2 Configuration point Selection/ Explanation setting option Washtimer (active) Activation/deactivation of the washtimer Washtimers are used to clean analysis sensors regularly and can actuate external equipment for cleaning sensors via digital outputs.
  • Page 152: Counters

    10 Configuration 10.15 Counters Open: Device menu r Configuration r Counter r Counter 1 to 4 Configuration point Selection/ Explanation setting option Function Inactive Counter operation mode Service count. Operation hours counter Service count.: The positive edges (switch-on operations) of a digital signal are counted (E.g.
  • Page 153: Ethernet

    JUMO AQUIS touch P is connected. To enter an IP configuration manually for the JUMO AQUIS touch P, a valid available IP address in the network must be known. Please contact your network administrator in this regard.
  • Page 154: Serial Interfaces

    Unambiguous specification of a bus participant 10.18 Formula Mathematical formula are created with the JUMO PC setup program and loaded to the device. Following this, the configuration of a mathematical for- mula can also be edited directly on the device.
  • Page 155: Logic Formula

    The unit set for the particular mathematical formula appears in the entry field. 10.19 Logic formula Logic formula are created with the JUMO PC setup program and loaded to the device. Following this, the configuration of a logic formula can also be ed- ited directly on the device.
  • Page 156: External Analog Inputs

    10 Configuration Configuration point Selection/ Explanation setting option -99999 to +99999 Start Upper/lower limit for labeling the scale Display range when displaying measuring values, e.g. in -99999 to +99999 recorder Display range diagrams and bar graphs The unit set for the particular manual value appears in the entry field. 10.21 External analog inputs Open: Device menu r Configuration r External Analog Inputs r External analog inputs 1 to 8...
  • Page 157: External Digital Inputs

    10 Configuration 10.22 External digital inputs Open: Device menu r Configuration r External Digital Inputs r External digital inputs 1 to 8 Configuration point Selection/ Explanation setting option Description up to 15 text characters Designation of the external digital input Hold value Nonvolatile storage of the last analog value re- ceived when the device restarts...
  • Page 158 10 Configuration Configuration point Selection/ Explanation setting option K-Faktor 0 to. 99999 l Ratio of the pulse count to flow rate (pulses per liter) The K-factor can be obtained from the documen- tation for the fitting in which the flow-through sen- sor (e.g.
  • Page 159 10 Configuration Configuration point Selection/ Explanation setting option Total quantity only available with flow rate measurement ac- hourly tivated: daily Activation of the flow-through quantity counter weekly monthly The setting establishes the automatic reset cy- annually cle. unlimited In addition, the flow-through quantity counters can be reset manually at the functional level or via a digital signal (reset input).
  • Page 160 10 Configuration...
  • Page 161: Retrofitting Optional Boards

    11 Retrofitting optional boards 11.1 Inserting optional boards DANGER! Insertion and removal of optional boards must be performed only by qualified personnel. To ensure electrical safety, country-specific regulations must be ob- served. The following step-by-step table explains in detail the procedure for retrofitting optional boards: Step Action...
  • Page 162 11 Retrofitting optional boards Step Action Disconnect all screw terminals and interface cables connected at the rear of the device. To prevent connections being interchanged, note down the assignment of the connectors to the sockets. Undo (do not remove) the two countersunk screws on the bottom of the case and remove the pan head screw on the side of the case.
  • Page 163 11 Retrofitting optional boards Step Action Insert the optional board in the selected slot. Ensure that the board is seated correctly. Fill all empty slots with plastic board frames. Re-insert the rear panel of the case and fasten with 3 screws. Make sure that all the toothed lock washers for securing the screws are re-inserted (see Step 7).
  • Page 164: Ci Base Calibration

    When the battery is empty a battery alarm will indicate that status. The battery must be exchanged before it is empty. The bat- tery must be exchanged by the JUMO Service department! In this case, send in the device!
  • Page 165 The factory settings authorize the "Master" and "Service" users for this. v Chapter 8.1.1 "Passwords and user rights", Page 65 Ensure that the electronics of the JUMO AQUIS touch P have reached operating temperature. You can view the board tempera- ture at: Device Menu r Service r Service Data r "Internal Data"...
  • Page 166 11 Retrofitting optional boards Step Action Place the wire of the calibration sensor with 2 windings through the opening in the Ci sensor without connecting the ends of the wires. Start the Ci base calibration Device Menu r Service r Ci Base Calibration IN 7 to 8 Enter the cell constant for the sensor and confirm by pressing "OK".
  • Page 167 11 Retrofitting optional boards Step Action Connect the ends of the wires forming the conductor loop of the calibration adapter. Set the calibration adapter to the resistance value shown in the in- struction text in the display (in the example: 20kΩ). Once the mea- surement displayed has stabilized, confirm by pressing "OK".
  • Page 168 11 Retrofitting optional boards Step Action Once all measurings have been confirmed, a summary of the cali- bration data acquired appears. Confirm by pressing "OK". If the Ci base calibration fails, the procedure is canceled without acceptance of the calibration data. Protocol after successful Ci base calibration...
  • Page 169: Calibration In General

    12 Calibration in general 12.1 Notes WARNING! During the calibration, the relays and analog output signals assume the config- ured states! 12.2 General information The actual electrical characteristics of analysis sensors always deviate some- what from the nominal specifications. The reasons for this include: •...
  • Page 170 12 Calibration in general • You must be logged in as a user with the right to perform calibrations. Fac- tory-preset users have all of these rights. v „Passwords and user rights“, Page 65 • You must ensure that the calibration default settings for the individual analy- sis inputs and, possibly, the universal inputs are set correctly.
  • Page 171: Calibration Logbook

    12 Calibration in general 12.3 Calibration logbook A separate logbook is maintained for each analysis and universal input. The last 10 successful calibrations of the input concerned are saved in the cal- ibration logbook. Canceled or failed calibrations (calibrations outside the per- missible limits) are not saved in the logbook, but rather noted in the event list.
  • Page 172 Tapping the "Details" button opens the selected logbook entry in the detail view. The detail view displays a table with all calibration values from a calibration pro- cedure. The "Service" button is used for diagnostic purposes by trained person- nel or JUMO Service.
  • Page 173 12 Calibration in general Assessment criteria pH calibrations (glass electrodes and ISFET connected to analysis measuring inputs as well as standard sig- nals connected to universal inputs) Calibration value [unit] — — < ≤ < < ≤ < Zero point [pH] 6 to 8 <...
  • Page 174 12 Calibration in general Calibration of conductivity sensors (analysis measuring inputs and standard signals connected to universal inputs) Calibration value [unit] — — < ≤ < < ≤ 150 < Rel. cell constant (CR) [%] 75 to 125 < ≤ <...
  • Page 175: Calibrating A Ph Measuring Chain

    13 Calibrating a pH measuring chain 13.1 Notes WARNING! During the calibration, the relays and analog output signals assume the config- ured states! 13.2 General information The calibration of pH electrodes is based on measurements in Buffer solutions with a defined pH-value. The pH values of the Buffer solutions used are speci- fied either via entry of fixed values into the calibration default settings, entered during the calibration or recognized automatically by "automatic buffer recogni- tion"...
  • Page 176: Calibration Default Settings For Ph Sensors

    13 Calibrating a pH measuring chain 13.2.2 Calibration default settings for pH sensors Before you can perform a calibration, you must first enter the necessary calibra- tion default settings. The possible settings for the pH Calibration are described in the following. Open the calibration default settings: Device Menu r Calibration r Select Analysis Input for pH/Redox/NH r Cali-...
  • Page 177 "pH buf- preferably for calibrating and fer 1 to 3" are hidden. adjusting technical pH mea- You need the JUMO PC setup program suring instruments acc. to DIN to edit Buffer set tables. 19267...
  • Page 178: Ph Calibration Routines

    13 Calibrating a pH measuring chain 13.3 pH Calibration routines NOTE! You must be logged in with corresponding user rights to perform calibrations. v Chapter 8.2.1 "Log-on/Log-out", Page 78 13.3.1 Zero-point calibration Step Action Start the Zero-point calibration. Device Menu r Calibration r Select analysis input for pH/Redox/ r Open Zero-point calibration If temperature compensation was not specified in the calibration default settings, enter the temperature of the Buffer solution here...
  • Page 179 13 Calibrating a pH measuring chain Step Action Entry of the pH-value of the Buffer solution • without buffer recognition: Check whether the "pH buffer 1" matches the pH-value of the Buffer solution used. If a Buffer set table was not specified, the "Buffer 1 pH"...
  • Page 180: Two-Point And Three-Point Calibration

    13 Calibrating a pH measuring chain 13.3.2 Two-point and Three-point calibration Step Action Start the desired Calibration routine. Device Menu r Calibration r analysis input Select analysis input for pH/Redox/NH3 r Open Two-point or Three-point calibration If temperature compensation was not specified in the calibration default settings, enter the temperatures of the Buffer solution here manually.
  • Page 181 13 Calibrating a pH measuring chain Step Action Entry of the pH-value of the Buffer solution • without buffer recognition: Check whether the "pH buffer 1" matches the pH-value of the Buffer solution used. If a Buffer set table was not specified, the "Buffer 1 pH"...
  • Page 182 13 Calibrating a pH measuring chain...
  • Page 183: Calibrating Redox Sensors

    14 Calibrating Redox sensors 14.1 Notes WARNING! During the calibration, the relays and analog output signals assume the config- ured states! 14.2 General information The calibration of Redox sensors is based on measurements in test solutions with a defined Redox potential. 14.2.1 Calibration methods for Redox sensors Zero-point calibration...
  • Page 184: Calibration Default Settings For Redox Sensors

    14 Calibrating Redox sensors 14.2.2 Calibration default settings for Redox sensors Before you can perform a calibration, you must first enter the necessary calibra- tion default settings. The possible settings for the Redox calibration are de- scribed in the following. Open the calibration default settings: Device Menu r Calibration r Select Analysis Input for pH/Redox/ r Calibration Default Settings...
  • Page 185: Redox Calibration Routines

    14 Calibrating Redox sensors 14.3 Redox calibration routines NOTE! You must be logged in with corresponding user rights to perform calibrations. v Chapter 8.2.1 "Log-on/Log-out", Page 78 14.3.1 Zero-point calibration Step Action Ensure that • the calibration default settings are correct, •...
  • Page 186: Two-Point Calibration

    14 Calibrating Redox sensors 14.3.2 Two-point Calibration Step Action Ensure that • the calibration default settings are correct • "Percent" is set as the Redox unit in the configuration of the Re- dox measuring input. v Chapter 14.2.2 "Calibration default settings for Redox sensors", Page 184.
  • Page 187: Calibrating Ammonia Sensors

    15 Calibrating ammonia sensors 15.1 Notes WARNING! During the calibration, the relays and analog output signals assume the config- ured states! 15.2 General information The calibration of ammonia sensors is based on measurements in ammonia- free test solutions. 15.2.1 Calibration methods for ammonia sensors Zero-point calibration This calibration method is used to determine the ammonia zero point.
  • Page 188: Ammonia Calibration Routines

    15 Calibrating ammonia sensors 15.3 Ammonia calibration routines NOTE! You must be logged in with corresponding user rights to perform calibrations. v Chapter 8.2.1 "Log-on/Log-out", Page 78 15.3.1 Zero-point calibration Step Action Start the Zero-point calibration. Device menu r Calibration r Select Analysis Input for pH/Redox/ r Zero-Point Calibration Clean the ammonia electrode and immerse it in the ammonia-free test solution.
  • Page 189: Calibrating Cr Conductivity Sensors

    16 Calibrating CR conductivity sensors 16.1 Notes WARNING! During the calibration, the relays and analog output signals assume the config- ured states! 16.2 General information The calibration of CR sensors is based on measurements in test solutions with a defined electrolytic conductivity. Since the electrolytic conductivity of liquids is temperature dependent, the temperature of the test solution must be sensed.
  • Page 190: Calibration Default Settings For Cr Conductivity Sensors

    16 Calibrating CR conductivity sensors 16.2.2 Calibration default settings for CR conductivity sensors Before you can perform a calibration, you must first enter the necessary calibra- tion default settings. The possible settings for the CR calibration are described in the following. Open the calibration default settings: Device menu r Calibration r Select CR Analysis Input r Calibration Default Settings...
  • Page 191 16 Calibrating CR conductivity sensors Calibration default settings for calibrating the relative cell constant Parameter Setting options Explanation Reference conductivi- 0 to 9999 mS/cm Conductivity of the reference solution Calibration default settings for calibrating the temperature coefficient Parameter Setting options Explanation Temperature com- Selection from...
  • Page 192: Cr Calibration Routines

    16 Calibrating CR conductivity sensors 16.3 CR calibration routines NOTE! Conductivity measuring inputs can be configured with measuring range change- over. Accordingly, calibrations must be performed for all "accessible measuring ranges". NOTE! You must be logged in with corresponding user rights to perform calibrations. v Chapter 8.2.1 "Log-on/Log-out", Page 78 16.3.1 Calibrating the relative cell constant...
  • Page 193 16 Calibrating CR conductivity sensors Step Action Ensure that • the sensor has been cleaned and is immersed in the test solu- tion, • the set reference conductivity matches the conductivity value of the test solution. Wait until the measuring value displayed stabilizes and then con- firm the result of the measurement by pressing "OK".
  • Page 194: Calibrating The Temperature Coefficient

    16 Calibrating CR conductivity sensors 16.3.2 Calibrating the temperature coefficient Step Action Start calibration of the temperature coefficient. Device menu r Calibration r Select CR Analysis Input r TC Cal- ibration Clean the sensor and immerse it in the sample solution. Ensure that the rel.
  • Page 195 16 Calibrating CR conductivity sensors Step Action • with temperature sensing A prerequisite is that temperature compensation was specified in the calibration default settings. Bring the temperature of the sample solution to the requested val- ues of the reference and operation temperatures in succession. The order does not matter.
  • Page 196 16 Calibrating CR conductivity sensors...
  • Page 197: Calibrating Ci Conductivity Sensors

    17 Calibrating Ci conductivity sensors 17.1 Important information WARNING! During the calibration, the relays and analog output signals assume the config- ured states! 17.2 General information The calibration of Ci sensors is based on measurements in test solutions with a defined electrolytic conductivity.
  • Page 198: Calibration Default Settings For Ci Conductivity Sensors

    17 Calibrating Ci conductivity sensors TC curve (for nonlinear temperature coefficients) If the conductivity of a liquid whose temperature coefficient changes with tem- perature has to be measured, this method can determine 5 temperature coeffi- cients for 5 temperature intervals. In this way, it is possible to determine a good approximation of the temperature coefficient curve.
  • Page 199 17 Calibrating Ci conductivity sensors Sample screen: Ci calibration default settings The calibration default settings enable the calibration routines to be accessed in the particular calibration menu. Calibration routines that are not enabled are not visible in the calibration menu. Additional calibration default settings are explained in the following table.
  • Page 200 17 Calibrating Ci conductivity sensors Calibration default settings for calibrating the TC curve Parameter Setting options Explanation Temperature com- Selection from Temperature input for automatic sensing of the test/ pensation analog selection sample solution temperature during the calibration Starting temperature -50 to 250 °C The starting and final temperatures of the range for which a temperature coefficient curve is to be deter-...
  • Page 201: Ci Calibration Routines

    17 Calibrating Ci conductivity sensors 17.3 Ci calibration routines NOTE! Conductivity measuring inputs can be configured with measuring range change- over. Accordingly, calibrations must be performed for all "accessible measuring ranges". NOTE! You must be logged in with corresponding user rights to perform calibrations. v Chapter 8.2.1 "Log-on/Log-out", Page 78 NOTE! Analysis inputs for inductive conductivity measurements (Ci) must undergo a...
  • Page 202 17 Calibrating Ci conductivity sensors Step Action Ensure that • the sensor has been cleaned and is immersed in the test solu- tion, • the set reference conductivity matches the conductivity value of the test solution. Wait until the measuring value displayed stabilizes and then con- firm the result of the measurement by pressing "OK".
  • Page 203: Calibrating The Temperature Coefficient

    17 Calibrating Ci conductivity sensors 17.3.2 Calibrating the temperature coefficient Step Action Start calibration of the temperature coefficient. Device Settings Menu r Calibration r Select Ci Analysis Input or Universal Input r TC Calibration Clean the sensor and immerse it in the sample solution. Ensure that the rel.
  • Page 204 17 Calibrating Ci conductivity sensors Step Action • with temperature sensing A prerequisite is that temperature compensation was specified in the calibration the default settings. Bring the temperature of the sample solution to the requested val- ues of the reference and operation temperatures in succession. The order does not matter.
  • Page 205: Calibrating The Tc-Curve

    17 Calibrating Ci conductivity sensors 17.3.3 Calibrating the TC-Curve Step Action Start the desired calibration of the TC-Curve. Device Menu r Calibration r Analysis Input 1 to 4 (Ci) and/or Universal Input 1 to 3 r TC-Curve Clean the sensor and immerse it in the sample solution. Ensure that the rel.
  • Page 206 17 Calibrating Ci conductivity sensors...
  • Page 207: Calibrating Universal Inputs

    18 Calibrating universal inputs 18.1 Important information WARNING! During the calibration, the relays and analog output signals assume the config- ured states! 18.2 General information 18.2.1 Calibration methods for universal inputs Universal inputs can be configured with various operation modes for a number of different process variables (see table below).
  • Page 208 18 Calibrating universal inputs Two-point calibration The zero point and slope of the measuring characteristic curve are determined with the aid of 2 measurings of 2 different reference solutions. Two test solutions with defined values of the respective process variable are needed as a references.
  • Page 209: Universal Inputs Calibration Default Settings

    18 Calibrating universal inputs 18.2.2 Universal inputs calibration default settings The calibration default setting available for selection depends on the configura- tion settings of the universal input. Open the calibration default settings: Device Menu r Calibration r Select Universal Input r Calibration Default Settings Calibration default settings for the individual operation modes •...
  • Page 210 18 Calibrating universal inputs • free chlorine, ph/Temp.-compensated In the universal inputs calibration default settings for the "free chlorine, ph/T- compensated" operation mode, the slope calibration is enabled and pre-config- ured as the only available calibration routine. Sample screen: Universal input calibration default settings in the "free chlorine, pH/T-com- pensated"...
  • Page 211: Universal Input Calibration Routines

    18 Calibrating universal inputs 18.3 Universal input calibration routines This chapter explains the universal inputs calibration routines for the "linear scal- ing" and "free chlorine, ph/T-compensated" operation modes. The explanations in the corresponding calibration chapters for the "pH value measurement" and "Conductivity measurement" operation modes apply, except that three-point calibration for pH sensors is not available for universal inputs (see Chapter 18.2.1 "Calibration methods for universal inputs", Page 207).
  • Page 212: Zero Point/Slope Calibration (Linear Scaling)

    18 Calibrating universal inputs 18.3.1 Zero point/slope calibration (linear scaling) Step Action Start the desired Calibration routine. Device Menu r Calibration r Select Universal Input r Zero-point calibration Clean the sensor and immerse it in the test solution. Wait until the measured value displayed stabilizes and then confirm the result of the measurement by pressing "OK".
  • Page 213: Two-Point Calibration (Linear Scaling)

    18 Calibrating universal inputs 18.3.2 Two-point calibration (linear scaling) Step Action Start the desired Calibration routine. Device Menu r Calibration r Select Universal Input r Two-point calibration Clean the sensor and immerse it in the first test solution. Wait until the measuring value displayed stabiliz- es and then confirm the result of the measurement by pressing "OK".
  • Page 214 18 Calibrating universal inputs...
  • Page 215: Slope Calibration (Free Chlorine, Ph/Temp.-Compensated)

    18 Calibrating universal inputs 18.3.3 Slope calibration (free chlorine, ph/Temp.-compensated) Step Action Start the slope calibration. Device Menu r Calibration r Select Universal Input r Slope Cal- ibration Clean the sensor and immerse it in the test solution. Check the displayed values for the pH value and temperature.
  • Page 216 18 Calibrating universal inputs Step Action Press "Yes" to accept the Calibration values determined and enter the Calibration in the Calibration logbook. Press "No" to discard the results.
  • Page 217: Pc Setup Program

    JUMO PC setup program. 19.1 General Information All configuration data and parameterization can be edited offline with the JUMO PC setup program. Entire device configurations can be loaded into the device, exported from the device and saved on a PC/laptop.
  • Page 218: Installing The Jumo Pc Setup Program

    DVD drive • USB host interface • LAN interface (Ethernet) NOTE! Close all applications on your PC before starting installation of the JUMO PC setup program. 19.2.1 Procedure Step Action With the PC running, insert the CD supplied with the device into the drive and close it.
  • Page 219 • Full version Insert the license key that you received from JUMO Select the program folder to which the links for starting the software art be copied. The directory for the program files is specified automatically.
  • Page 220: Setup Program Login

    19 PC Setup Program To install the full version, you must acquire an appropriate license from JUMO. The contact data can be found on the back of this document. NOTE! If the "30-days test version" option is selected during installation, the JUMO PC setup program is fully functional for 30 days.
  • Page 221: Rights In The Setup Program

    19 PC Setup Program 19.3.1 Rights in the setup program Depending on the version and login, individual users have different rights in the PC setup program. The differences are summarized in the following table. Demo Service Specialist Users Version User rights Open Save, Save As, Delete...
  • Page 222: Quickstart Agent

    19 PC Setup Program 19.4 QuickStart agent Installation During installation of the PC setup program, an additional program with the de- scription "QuickStart agent" is installed. This program starts automatically and appears as a symbol in the Windows taskbar. Example: "Show icons"...
  • Page 223 If autostart is inactive, i.e. the symbol does not appear in the taskbar, the Quick- Start agent can be started manually from the Windows Start menu. The Quick- Start agent is located in the same program group as the JUMO PC setup program.
  • Page 224: User Interface

    19 PC Setup Program 19.5 User interface 19.5.1 Elements of the user interface Menu bar Toolbar Working area – navigation tree Working area – display window Connection status Online data Menu bar The individual functions of the PC setup program are launched from the menu bar.
  • Page 225 19 PC Setup Program Moving the toolbar The user can change the position of the toolbar: Step Action Position the mouse pointer between two icon groups Left-click with the mouse button While holding down the left mouse button, move the toolbar to the desired location Release the left mouse button Possible positions of the toolbar:...
  • Page 226: Display Protection

    19 PC Setup Program Online data The "Online data" function displays current process data in the PC setup pro- gram. The "Window" menu can be used to show or hide the online data window. Window r Online Data The height of the window can be changed by moving the border of the working area or the "Connection status"...
  • Page 227 19 PC Setup Program When display protection is activated, all device settings in the PC setup program are concealed the next time the setup file is opened. To show the device settings in the PC setup program again, double-click the en- try "Display protection"...
  • Page 228 19 PC Setup Program...
  • Page 229: Editing The Setup File

    Open an existing file: File r Open The working area (navigation tree and display window) shows the menu struc- ture with the individual setting options for the JUMO AQUIS touch P from within the JUMO PC setup program. Navigation tree...
  • Page 230 19 PC Setup Program Navigation tree Left-clicking (once) on the entry "Project" or on a certain module displays the as- sociated subitems in the display window. Clicking on the ("-") symbol closes a menu, clicking on the ("+") symbol opens a menu. Double-clicking on an entry (e.g.
  • Page 231: Menu Bar

    19 PC Setup Program 19.6 Menu bar This chapter describes the functions of the menu bar. The order of the subchap- ters corresponds to the location of the menus in the menu bar (from left to right). 19.6.1 File menu Creates a new setup file in the working area.
  • Page 232 19 PC Setup Program Page View ... The printed result is displayed on the screen. Several pages can be displayed and the size of what is displayed changed. Printer Setup ... Permits the printer settings to be changed. When a program starts, the Windows default printer is always used as the active printer.
  • Page 233: Data Transfer Menu

    19 PC Setup Program 19.6.2 Data Transfer menu Establish Connection ... Opens the device connections list. The content of the device connections list is project-dependent. The device connections list contains all devices to which a connection can be established via the setup program. Devices can be added to or removed from the list.
  • Page 234 Imports a setup file in SET format from external mass storage (USB flash drive) into an open setup file. In this way, an SET file that was previously copied from the device to a USB flash drive can be opened for editing with the JUMO PC set- up program.
  • Page 235: Extras

    (demo mode). Renew login / Change password Opens a window to activate user login. After the JUMO PC setup program has been installed, prompts for the user name and password do not appear until user login has been activated.
  • Page 236 19 PC Setup Program Version library Shows the content of the version library. The list contains the standard versions of the device software and the compatible hardware versions provided by the device manufacturer together with the PC setup program. Additional software versions that were imported subsequently (e.g.
  • Page 237: Windows

    19 PC Setup Program 19.6.4 Windows Cascaded Arranges all open setup windows in an overlapping arrangement. Left-clicking on a window brings it to the foreground. Tiled Arranges all open setup windows in a tiled arrangement. Left-clicking on a win- dow makes it the active window. Arrange icons Arranges the symbols for all minimized setup windows in the lower region of the user interface.
  • Page 238: Info

    19 PC Setup Program 19.6.5 Info Setup info Displays information about the setup program, incl. the version number. The version number is important, for instance, when contacting the service hotline. Registered license numbers ... Shows all registered license numbers and any active options. The license num- bers and the information about active options are important, for instance, when contacting the service hotline.
  • Page 239: Connection To Device

    USB connector cannot be prevented due to the lack of a captive connection. An existing connection is a requirement for data transfer between the JUMO PC setup program and system. In the menu bar, open: Data Transfer r Establish Connection ...
  • Page 240 19 PC Setup Program there is no connection), the PC setup program opens the device connection list. The user must first highlight the device concerned in the device connection list and then press the "Establish connection" button. If a device should not be con- nected, the corresponding dialog ("Device connection list"...
  • Page 241: Configuring The Connection With The Assistant

    19 PC Setup Program 19.7.2 Configuring the connection with the assistant The connection settings assistant assists the user with configuring a new con- nection. The assistant is always active after opening the device connections list (check mark set). It is started whenever a new entry is to be added to the device connection list.
  • Page 242 19 PC Setup Program Setup dialog - Device and type of connection Adjustments Setting item Selection/ Explanation setting option Device version Selection from Device version of the device to be con- dropdown list nected v Chapter 8.2.4 "Device information", Page 82 Connection pre- Setup or ring mem- LAN interface (HTTP protocol), connec-...
  • Page 243 19 PC Setup Program Setup dialog - Device login Adjustments Setting item Selection/ Explanation setting option Checkboxes Activate option (set checkmark) if connec- Do not log in tion to the device takes place without login. Without logging in, some functions probably will not be available.
  • Page 244 19 PC Setup Program Setup dialog - TCP/IP port settings This dialog depends on the selected connection preset. The figure shows the case "Setup or ring buffer via HTTP proxy". Except for the parameters "Proxy" and "Proxy Port", the description applies also for the case "Setup or ring buffer via TCP/IP".
  • Page 245 19 PC Setup Program After the user clicks on the "Next" button, the assistant opens the next setup di- alog. Setup dialog - Overview of the settings To complete the configuration, the user is presented with an overview of the set- tings.
  • Page 246 19 PC Setup Program Setup dialog - USB interface settings on the PC If several devices are connected via a USB interface, one device must be se- lected here. Furthermore, various test criteria can be activated so that the con- nection to the device is established only after a successful test.
  • Page 247 19 PC Setup Program To complete the configuration, the user is presented with an overview of the set- tings. An incorrect setting can be corrected by pressing the "Exit" button to re- open the setup dialog concerned and making the correction there. Clicking on the "Finish"...
  • Page 248: Configuring The Connection Without The Assistant

    19 PC Setup Program 19.7.3 Configuring the connection without the assistant The assistant can be deactivated in the device connections list (remove the checkmark). v Chapter 19.7.1 "Device connections list", Page 239 During configuration, the user is not guided to the next step - as with an active assistant - but instead must select the individual steps of the configuration on their own.
  • Page 249 19 PC Setup Program Step Action Make the protocol and interface settings and click on OK. If neces- sary, click on the "Extended" button beforehand to make additional settings for communication. Close all previously opened setup dialogs (windows) by clicking on the "OK"...
  • Page 250: Searching For A Device In The Network

    19 PC Setup Program 19.7.4 Searching for a device in the network This function assists the user with selecting the IP address or the host name for the device of interest. A "Search" button is located to the right of the entry field in the dialog for entering the IP address or host name.
  • Page 251 19 PC Setup Program enters the host name (DNS name) in the input field. You can use the "Check and convert host name" function to establish the IP ad- dress and transfer it to the input field (subsequent editing may be necessary in the input field).
  • Page 252: Setting Up The Device With The Jumo Pc Setup Program

    Chapter 9 "Parameterization", Page 109 19.8.2 User list The factory-entered user names, passwords, and user rights (see Chapter 8.1.1 "Passwords and user rights", Page 65) can be changed with the JUMO PC setup program. Configuration point Selection/ Explanation setting option...
  • Page 253 19 PC Setup Program Character table for passwords Characters for user login can be added or changed in character tables 5 to 16. Character tables 1 to 4 cannot be changed. Classification selection field Visualization of the keyboard layout in the Device menu Entry field for character set Header character set for labeling the tab in the device settings menu Table 1 to 16 selection field...
  • Page 254 19 PC Setup Program Configuring public Rights The access rights for all users who are not logged in are set here. Open Settings: Only Setup r User list r Public rights...
  • Page 255: Country Settings

    19 PC Setup Program Configuring user rights The user rights for each individual user account are set here. A corresponding user login is required to exercise the respective user rights. v Chapter 8.1.1 "Passwords and user rights", Page 65 v Chapter 8.2.1 "Log-on/Log-out", Page 78 Open Settings: Only Setup r User list r Rights 19.8.3...
  • Page 256 19 PC Setup Program Setting item Selection/ Explanation setting option No entry possible Specification Internal Specification (only for Assigned automatically service purposes to identify the language) Language Description of the language Use factory-set text or enter a different text Date format DDMMYYYY Display format for the date MMDDYYYY...
  • Page 257 19 PC Setup Program Language button Languages can be created, edited, and deleted with the "Language" button. New languages can be created on the basis of already existing languages. Available languages as well as languages from libraries can be selected as a template for new languages in the open setup.
  • Page 258 19 PC Setup Program From a language in the current library (internal library of the JUMO PC set- up program) Step Action Open Country Settings: Only Setup (double-click) r Country Settings (double-click) Open the current library: Button "Language“ r New Language r From Current Library...
  • Page 259 Editing languages NOTE! Standard languages cannot be edited. Only languages created by the user can be edited with the JUMO PC setup program. v Chapter "Creating a new language", Page 257 The following figure shows the source language text on the left and the target language text on the right.
  • Page 260 19 PC Setup Program Toolbar The toolbar gives the user access to various functions for editing text. Explanation Button Search text Replace text with other text Use filter to reduce the number of texts in the list Sort texts alphabetically Display text ID (instead of Row number) Mark text as "not edited closed votes (red text) Mark text as "to be checked"...
  • Page 261: Device Character Set

    19 PC Setup Program 19.8.4 Device character set When creating a new language, it is also possible to specify the characters that appear on the device for text entry (virtual keyboard). Entry field for character selection The characters that are entered here are offered for selection by the vir- tual keyboard on the device later upon pressing the corresponding key.
  • Page 262 19 PC Setup Program To edit the character set for the device keyboard, proceed as follows: Step Action Left-click the virtual key in the "Character entry" field (example: "+"). The characters assigned to the key appear in the text field. The first character from the left corresponds to the key label.
  • Page 263: User Level

    19 PC Setup Program 19.8.5 User level The user-defined parameter and settings list at the User level of the Device menu are created in this menu. Explanation of the User level: v Chapter 8.2.2 "User level", Page 79 Open Settings: Only Setup r User level...
  • Page 264: Formula

    Formula must be created with the JUMO PC setup program. During data trans- fer to the device, the formula are then loaded into the JUMO AQUIS touch P to- gether with the setup. For formula that have already been loaded into the device, the configuration data below the formula entry field can also be set on the device as an alternative.
  • Page 265: Logic Formula

    Logic formula must be created with the JUMO PC setup program. During data transfer to the device, the formula are then loaded into the JUMO AQUIS touch P together with the setup. For formula that have already been loaded into the device, the description below the formula entry field and the alarm configuration can also be set on the device as an alternative.
  • Page 266: Customer-Specific Linearization

    JUMO PC setup program. During data transfer to the device, the linear- ization is then loaded into the JUMO AQUIS touch P together with the setup. Af- ter the linearizations have been loaded into the device, they can be used when configuring the following types of analog inputs: •...
  • Page 267 19 PC Setup Program Symbol buttons – Basic values table "Insert line" Inserts a new line above the highlighted line "Delete line" Deletes the highlighted line from the Basic values table "Delete table" Delete all entries in the Basic values table "Generate formula"...
  • Page 268 19 PC Setup Program Settings for customer-specific linearizations 1 to 8 Setting item Selection/ Explanation setting option Description up to 15 text characters Linearization name Linearization type Basic values Way of entering linearization data Formula In order to display a measurement characteristic curve, the customer-specific linearization can be entered as a value table with discrete value pairs or as a formula...
  • Page 269: Buffer Set Tables

    19 PC Setup Program 19.8.9 Buffer set tables Three Buffer set tables are created in the JUMO AQUIS touch P. Buffer set tables are needed for calibration of pH sensors when using automatic buffer detection. v Chapter 13 "Calibrating a pH measuring chain", Page 175 These Buffer set tables can be changed as often as necessary here.
  • Page 270: Process Screens

    19.8.10 Process screens A "Process screen" can be created in the Operation loop of the JUMO AQUIS touch P in order to design a dynamic overview display of a system or a process. The Process screen must be created with the JUMO PC setup program and loaded into the JUMO AQUIS touch P together with the configuration.
  • Page 271 19 PC Setup Program Process screen editor Process screens for the JUMO AQUIS touch P must be created with the JUMO PC setup program. These are then transferred from the PC to the device togeth- er with the setup. Description of the Process screen (appears in the title bar of the Process screen)
  • Page 272 19 PC Setup Program Process screen Editor buttons "Cut" Copies the highlighted object to the clipboard and deletes it from the list "Copy" Copies the highlighted object to the clipboard "Insert" Inserts an object from the clipboard at the highlighted posi- tion in the list;...
  • Page 273 19 PC Setup Program Preview The "Preview" button in the Process screen Editor opens the preview. The con- cept for the Process screen can be viewed here. The Process screen is dis- played as it will appear later on the device. Objects can be repositioned via drag and drop.
  • Page 274 19 PC Setup Program "Move" Opens a window with arrow buttons from moving objects one step at a time Single arrow: Move by 1 pixel Double arrow: Move by 10 pixels "Align left" Simultaneously aligns several highlighted objects along the left edge of the object highlighted first "Align right"...
  • Page 275 19 PC Setup Program Background The next level in front of the background color level is reserved for the back- ground. Any image in the image list can be selected and displayed here in front of the background color level. User-defined images can be imported into the image list.
  • Page 276 19 PC Setup Program Importing/exporting images Step Action Open the image list Only Setup r Process screens r "Process screen 1" tab r "Background" button" r "Image List" button Import: Highlight an image in the list that you wish to overwrite with the new one and click the "Replace"...
  • Page 277 19 PC Setup Program Objects The object list of the Process screen Editor contains 50 empty objects. Double- clicking on an entry or clicking the "Edit" button in the Process screen Editor opens a dialog for configuring, designing, and formatting the corresponding ob- ject in detail.
  • Page 278 19 PC Setup Program Analog signals: The numerical value of an analog signal is displayed in the Process screen. Configuration item Selection/ Explanation setting option Visible Checkmark in the Checkmark set: Object is displayed checkbox Empty checkbox: Object is hidden X position 0 to 315 pixels X-coordinate of the top left corner of the picto-...
  • Page 279 19 PC Setup Program Digital signals: A binary signal has two plain texts associated with it. Depending on the value of the binary signal, one of the two plain texts is displayed. Example: A manual / automatic selector switch needs to be visualized. Depend- ing on the switch position, either "Manual mode"...
  • Page 280 19 PC Setup Program Universal display: Display of configuration settings, e.g. units or identifiers of an internal analog signal source for the JUMO AQUIS touch P. Configuration item Selection/ Explanation setting option Visible Checkmark in the Checkmark set: Object is displayed...
  • Page 281 19 PC Setup Program Text: Plain text for labeling of visualization elements. Configuration item Selection/ Explanation setting option Visible Checkmark in the Checkmark set: Object is displayed checkbox Empty checkbox: Object is hidden X position 0 to 315 pixels X-coordinate of the top left corner of the picto- gram Y position 0 to 181 pixels...
  • Page 282 19 PC Setup Program Configuration item Selection/ Explanation setting option Color Selection from For rectangles only: color palette Fill color of the rectangle (transparent for objects) Frame type None Selection of a frame format Fine Bold Upper case Lower case Bar graph: Vertical or horizontal bar for visualizing analog values.
  • Page 283 19 PC Setup Program Example: Objects now need to be added to the sample Process screen from the previous chapter (Chapter "Background", Page 275). Valves and pumps are represented by pictograms. Two analog displays serve as the process value dis- play.
  • Page 284: Email

    The JUMO AQUIS touch P can be configured for event-controlled sending of e- mail notifications. Up to 5 e-mail templates can be stored in the JUMO AQUIS touch P. Sending of the e-mail is triggered by a rising edge of the binary signal in the "Alarm signal"...
  • Page 285: Sms Gateway

    19 PC Setup Program SMTP server: serves to transmit e-mails. In many networks, the authorization to send e-mails via a mail account must be released by previously logging in at the correspond- ing POP3 server. Proxy: serves as a gateway from the local company network to the worldwide Internet. This is also where the conversion of "local"...
  • Page 286 19 PC Setup Program Setting up e-mail templates Open Settings: Only Setup r E-mail (double-click) r "E-mail 1 to 5" tab The following table explains the settings for event-controlled sending of e-mails. Configuration point Selection/ Explanation setting option E-mail addresses Enter a valid e-mail ad- Up to 3 e-mail addresses to each of which a par- 1 to 3...
  • Page 287 19 PC Setup Program Setting up the e-mail server To send messages, you must specify the data for the outgoing mail server. Open the server settings window and enter the information here. You can obtain the server settings from your e-mail service provider or network administrator. Open the dialog for the server settings: Only Setup r E-mail (double-click) r "E-mail Server"...
  • Page 288 19 PC Setup Program NOTE! An e-mail server within the company network must be entered (not on the In- ternet or accessible without proxy addressing)! This mail server should be able to also forward e-mails to the Internet.
  • Page 289 19 PC Setup Program Procedure for sending an e-mail over the Internet To better understand the process of sending an e-mail and correct any errors in the configuration of e-mail servers and Ethernet, a graphical representation of the processes is presented in the following. Conditions fulfilled to send an e-mail? Loading of current...
  • Page 290 19 PC Setup Program Checking the e-mail function Proper execution of the e-mail sending process can be checked in the event list. When an e-mail is sent successfully, a corresponding entry is made in the event list. Entries are made in the event list even in the case of a fault and can be help- ful when diagnosing a problem.
  • Page 291: Web Server

    NOTE! Creating websites requires the appropriate knowledge about website programming and specifying the webserver In the JUMO AQUIS touch Customer-specific websites can be imported into the configuration of the web- server. To edit an already saved website, it can be exported. Data from the...
  • Page 292 19 PC Setup Program Open Settings: Only Setup r Configuration Webserver (double-click) (2) (3) "Import Web" button "Export Web" button "Delete Web" button Window with tree view of the saved website "HTML Tags" button Opens a dialog for copying the IDs of internal data from the device memory These IDs are needed to request the internal data from the web document.
  • Page 293: Calibration Logbook

    19.8.14 Calibration logbook As an alternative, the calibration logbooks for the analysis sensors can also be viewed in the JUMO PC setup program. Data regarding calibrations of analysis sensors are kept in the calibration logbooks. v "Calibration logbook", page 171 Open: Only Setup r Calib.
  • Page 294: Online Parameters

    The device must thus be connected via a USB or Ethernet port to the PC on which the JUMO PC setup program is installed, and a connection be- tween the JUMO PC setup program and device must be established.
  • Page 295: Screenshot

    The Ethernet settings are explained in the configuration. v See chapter 10.16 "Ethernet", page 153 NOTE! The Ethernet settings are entered in the online parameters in the JUMO PC setup program. The IP configuration of the device can also be changed from the PC if the PC and device are connected via Ethernet.
  • Page 296: Enabling Of Extra Codes

    • Enter activation code After you have sent the code number for the desired extra code to JUMO, you receive the activation code for this option. Following correct entry of the activation code, this option is available for use in the device.
  • Page 297: Adjusting / Testing

    When "Adjusting / Testing" is opened, the outputs assume undefined states. For this reason, "Adjusting / Testing" may not be opened during operation of equipment being controlled by the JUMO AQUIS touch P. Numerous testing functions that are available in similar form in the device set- tings menu at the "Functional level"...
  • Page 298 19 PC Setup Program Analog inputs Proper operation of the analog inputs of the device can be checked here. In addition to selecting the analog input and signal type, the line frequency of the power supplied to the device must be set in order to test an analog input correct- ly.
  • Page 299 19 PC Setup Program Analog outputs Proper operation of the analog outputs of the device can be checked here. A desired output value can be specified in the "Setpoint value" field for the se- lected output and signal type set. Pressing the "Test" button starts the test. During the test, a window for monitoring the output signal is displayed.
  • Page 300 19 PC Setup Program Binary outputs Proper operation of the Digital outputs of the device can be checked here. After setting checkmarks in the option fields, the Digital outputs of the device can be controlled manually via the PC setup program. The "Set All" and "Delete All" buttons allow all Digital outputs to be switched on and off at once.
  • Page 301 19 PC Setup Program Binary inputs Proper operation of the Digital inputs of the device can be checked here. The "Read Continuously" button starts a continuous test that can be stopped by pressing the "Stop" button. During this test, the binary values of the device in- puts are read continuously and displayed in acc.
  • Page 302 Touch In this view, touchscreen calibration (see "Calibrating the touchscreen", page 84) can be started from the JUMO PC setup program. In addition, the calibration data for the touchscreen are displayed in detail. These are used to diagnose problems with touch operation together with JUMO Service.
  • Page 303: Technical Data

    20 Technical data 20.1 Analog inputs base unit 20.1.1 Temperature measuring input (IN 4) Probe-/Signal type Connection type Measuring Measuring accu- Ambient range racy temperature influence -200 to +850 °C ≤ 0.05% of MR ≤ 50 ppm/K Pt100 DIN EN 60751 2-wire/3-wire -200 to +850 °C ≤...
  • Page 304: Temperature Measuring Input (In 5)

    20 Technical data 20.1.2 Temperature measuring input (IN 5) Probe-/Signal type Connection type Measuring Measuring accu- Ambient range racy temperature influence -200 to +850 °C ≤ 0.05% of MR ≤ 50 ppm/K Pt100 DIN EN 60751 2-wire/3-wire -200 to +850 °C ≤ 0.1% of MR ≤...
  • Page 305: Analog Inputs Optional Boards

    20 Technical data 20.2 Analog inputs optional boards 20.2.1 Universal input (IN 11, IN 12) Probe-/Signal type Connection type Measuring Measuring accu- Ambient range racy temperature influence -200 to +850 °C ≤ 0.05% of MR ≤ 50 ppm/K Pt100 DIN EN 60751 2-wire/3-wire -200 to +850 °C ≤...
  • Page 306: Analysis Input: Cr (Resistive Conductivity)

    20 Technical data 20.2.3 Analysis input: CR (resistive conductivity) Units µs/cm ms/cm kΩ × cm MΩ × cm Display ranges 0.0000 to 9.9999 00.000 to 99.999 000.00 to 999.99 0000.0 to 9999.9 00000 to 99999 Temperature compensation TC-linear, natural water DIN EN 27888, natural water with expanded range, ASTM D-1125-95 for neutral (NaCl), acidic (HCl) and alkaline (NaOH) impurities...
  • Page 307: Analysis Input: Ci (Conductivity, Inductive)

    20 Technical data 20.2.4 Analysis input: Ci (conductivity, inductive) Units µs/cm ms/cm Measuring/display ranges 0.0000 to 9.9999 00.000 to 99.999 000.00 to 999.99 0000.0 to 9999.9 00000 to 99999 Temperature compensation TC linear TC curve natural water natural water with expanded temperature range NaOH 0 to 12% NaOH 25 to 50% 0 to 25%...
  • Page 308: Temperature Compensations

    20 Technical data 20.2.5 Temperature compensations Compensation type Compensation range TC linear -50 to +250 °C TC curve -50 to +250 °C -50 to +250 °C natural water according to DIN EN 27888 0 to 36 °C natural water with 0 to 100 °C expanded temperature range ASTM D-1125-95 (neutral, alkaline, and...
  • Page 309: Measuring Circuit Monitoring, Optional Boards

    20 Technical data 20.2.6 Measuring circuit monitoring, optional boards Input/Sensor Underrange/ Short circuit/ Open circuit Special fea- overrange sensor break tures pH-value (glass elec- Configurable Configurable trode) Impedance mea- Impedance mea- surement surement pH-value (ISFET) Resistive conductivity Configurable Only with 4-wire circuit Inductive conductivity Universal input for con-...
  • Page 310: Analog Outputs Of Base Unit And Optional Boards

    This specification is relevant only if the Ext. voltage supply is selected under the “Contact” point in the Configuration. Sensors and transmitters should be supplied from voltage supply outputs on the JUMO AQUIS touch P. An externally supplied voltage signal must not have a voltage over 28 V.
  • Page 311: Binary Outputs, Optional Boards

    20 Technical data 20.7 Binary outputs, optional boards Optional card Switching out- Ampacity at re- Contact life Special features sistive load Relay output, dual nor- 2 normally open 3 A at AC 250 V 150,000 mally open contacts contacts switching cycles Relay output, single 1 changeover changeover contact...
  • Page 312: Interfaces

    20 Technical data 20.9 Interfaces 20.9.1 Serial interface RS485 (base unit ) Protocol Data formats Device address- Baud rates in Connection baud Modbus (slave) 8 - 1 - no parity 1 to 254 9600 Spring-cage 8 - 1 - odd parity 19200 terminals 8 - 1 - even parity...
  • Page 313: Ethernet Optional Board (10/100Base-T)

    Modbus TCP/IP enables Modbus users to communicate via a LAN, provided this is connected to the LAN (e.g. via gateways). To configure Modbus communication you will require the interface description for the JUMO AQUIS touch P. Enlist the help of your network administrator or an IT specialist for the IP-configuration.
  • Page 314: Usb Interfaces

    USB 2.0 data memory Type A Read/write device settings, Save Service data Update the firmware USB device interface Device setting via PC JUMO PC setup USB port setup program, program, Type Mini-B Extract, archive, evalu- JUMO PCC/PCA 3000 measurement data software The recording function stores measurement data in a ring buffer inside the device.
  • Page 315: Electrical Data

    20 Technical data 20.10 Electrical data Voltage supply (switch-mode) AC 110 to 240 V +10/-15%; 48 to 63 Hz or AC/DC 24 V +30/-25 %; 48 to 63 Hz electrical safety According to EN 61010, part 1 overvoltage category III, pollution degree 2 Max.
  • Page 316: Case

    20 Technical data 20.12 Case Site altitude maximum 2000 m above sea level Case type Plastic front frame with metal case barrel (for indoor use only) Materials Plastic front panel UL 94 V0 Case barrel from zinc-plated steel sheet Dimensions of the front 96 mm ×...
  • Page 317: Functions

    20 Technical data 20.13 Functions 20.13.1 Controller channels Number Controller type Two-state controllers Three-state controllers Continuous controller Coarse/fine controller Modulating controllers Continuous controller with position controller Controller structure P, PI, PD, PID Controller outputs For each controller channel, 2 outputs configurable as: pulse length output, pulse frequency output, (maximal 240 pulses per minute), continuous output Disturbance feedforward con-...
  • Page 318: Recording Function

    History function Archiving/evaluation Yes, (with JUMO PCA 3000 evalu- ation software ) A freely configurable set of input variables can be pooled in one group. Each group has its own display screen. The group affiliation is considered for data storage to enable eval- uation via PC.
  • Page 319 20 Technical data...
  • Page 320 20 Technical data...
  • Page 322 JUMO GmbH & Co. KG JUMO Instrument Co. Ltd. JUMO Process Control, Inc. Street address: JUMO House 6733 Myers Road Moritz-Juchheim-Straße 1 Temple Bank, Riverway East Syracuse, NY 13057, USA 36039 Fulda, Germany Harlow - Essex CM20 2DY, UK Phone:...

Table of Contents