16 Calibrating Ci conductivity sensors
48
Step
Action
3
Ensure that
•
the sensor has been cleaned and is immersed in the test solu-
tion,
•
the set reference conductivity matches the conductivity value of
the test solution.
Wait until the measuring value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK".
The preset reference conductivity can be changed here manually
if necessary.
tap button to change
reference range
4
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
5
Press "Yes" to accept the calibration values determined and enter
the calibration in the calibration logbook.
Press "No" to discard the results.