Power up the MAMMOMAT Stand

Radiation protection door switch (optional)

It is optional to connect a radiation protection door switch to the MAMMOMAT Novation
master board D750 connector X754.
NOTE
A 15V / 5 mA signal is used in the door switch circuit. When the signal path is opened by
the door switch, the exposure will immediately be terminated.
Fig. 1 Connecting the door switch
1.
If a radiation protections door switch is not used, a jumper, see Fig. 1, on the master board
750 at connector X754 (Pin 1,2) has to be inserted.

External exposure lamp (optional)

This should only be performed if the customer requires it.
The connection on the Master board works as a shutter. The signal from the Master board
is only active during exposure, thus the exposure lamp will only be lit when the signal is
active.
Connect the cable to the lamp holder in series with the power supply.
CAUTION
Siemens AG
Medical Solutions
A radiation protection door switch is not delivered with the MAM-
MOMAT Novation
To connect a door switch, remove the jumper from the X754 (Pin 1,2) located on
the master board D750, see Fig. 1, and connect the door switch instead.
When connecting the exposure lamp to the Master board, make
sure that the electrical load does not exceed 24VAC / 25W
DR
.
SPB7-250.815.02
Rev. 02
09.04
3
MAMMOMAT Novation
Page 1 of 6