9.6.3

Importing an STL file

1. Create a case in the ADMINISTRATION phase.
2. Then continue to the SCAN phase.
Ä The jaw of the preparation to be scanned is pre-selected in the
object list.
3. In the object list, select "Import STL" as the method. Open an STL
file via the file path.
4. As soon as the STL file is loaded, continue to the MODEL phase.
9.6.4
Scanbody scans for scanning with inPost and
ATLANTIS-FLO-S scanbodies
Use scanbody scans for the measurement of one or more implant
positions with the scanbody inPost for directly screwed on bridges on
multi-unit abutments.
CAUTION
Pay attention to the inPost scanbodies / ATLANTIS-FLO-S
scanbodies / multi-unit abutments matching the implant system.
Ensure the correct seating on the laboratory analog of the screwed in
scanbodies and multi-unit abutments prior to scanning. Elements
screwed in incorrectly may lead to the incorrect detection of the
implant position and thus to inappropriate restorations.
Observe the operating instructions for the scanbodies.
When screwing in the Sirona in-Post scanbodies for multi-unit
abutments, you must use commercially available latex gloves.
The ATLANTIS-FLO-S scanbodies can be screwed in without the use
of gloves.
1. Select the corresponding Scanbody image catalog in order to be
able to record the implant location for directly screwed bridges in
connection with the inPost or FLO-S scanbodies.
2. Clamp the model without screwed in scanbodies and gingival mask
to the arm of the scanner and start the overview scan.
Ä The "Capture Scanbody" button becomes active once the
overview scans are complete.
3. Click on the "Capture Scanbody" button.
NOTICE
The scanner moves into an appropriately easy-to-access position for
screwing in the scanbodies.
➢ Do not take the model down from the holder.
4. Screw the scanbody onto the first implant position.
5. Rotate the model in the 3D preview in the implant axis and double-
click on the position in which you screwed the scanbody in
beforehand.
9 Performing 3D acquisition
9.6 Example applications
53