Switching The Unit On And Off - Dentsply Sirona CEREC Primemill Operating Instructions Manual

Hide thumbs Also See for CEREC Primemill:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions CEREC Primemill
67 19 681 D3439
D3439.201.01.01.02    09.2019
NOTE
Damage to surfaces!
In the undiluted state, the coolant additive DENTATEC disintegrates
plastic surfaces and can cause discoloration.
➢ Do not place DENTATEC on the unit.
➢ Do not spill DENTATEC.
4. Add approx. 75 ml of DENTATEC to the tank.
5. Fill the tank with water until the filter insert is completely immersed
(up to the bottom edge of the cover thread, approx. 3.5 liters).
6. Wait for a short time until the filter insert is completely soaked; then
add an appropriate amount of water.
7. Close the water tank by tightening the tank cap clockwise by hand.
Do not use the tank cap opener for this.
8. Push the water tank back into the housing.
9. Close the drawer of the unit.
10. Select the symbol shown on the left in the touch display.
11. Switch the water pump on to fill the water circuit.
12. Leave the water pump running until a constant jet of water hits the
instruments (approx. 10 seconds).
13. Switch the water pump component off.
14. Fill the water tank up again until the filter insert is completely
immersed (up to the bottom edge of the cap thread).
15. Select the symbol shown on the left in the touch display.
16. To reset the water tank counter, press the "Check" button next to
the "Water Tank" category.
5.4.8

Switching the unit ON and OFF

NOTE
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment,
condensation may form and result in a short circuit.
Within the machine, grease depots are included for lubricating
components that can cause error messages at low temperatures.
ü Install the unit at room temperature.
➢ Wait until the unit has reached room temperature and is
absolutely dry (for at least one hour)
Ä The unit is dry and can be put into operation.
Switch on the unit
ü The unit is connected to the power supply.
1. The main switch on the rear side of the unit is set to position I (ON).
The On/Off button lights up blue.
2. Press the On/Off button on the front.
Ä The unit switches on and the On/Off button changes from blue
to orange.
3. Select the appropriate language and region.
Ä The production unit is looking for hub in the network.
5 Installation and startup
5.4 Commissioning
35