Evolution® Shortie RL Controlled-Rotation Dilator Sheath Set
Federal law restricts this device to sale by or on the order of a physician.

DEVICE DESCRIPTION

The Evolution® Shortie RL Controlled-Rotation Dilator Sheath Set is comprised of an inner polymer sheath
connected to a handle capable of mechanically rotating the sheath and an outer telescoping polymer
sheath. The inner sheath has a stainless steel tip at its distal end.

INTENDED USE

Evolution® Shortie RL Controlled-Rotation Dilator Sheath Set is intended for use in patients requiring the
percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects.
Sheath Size
9 Fr
11 Fr
Use With Other Devices
The Evolution® Shortie RL Controlled-Rotation Dilator Sheath Set may be used in conjunction with the
following catheter/lead extraction devices from Cook:
Locking Stylet
Lead Extender
Be sure to closely follow the "Suggested Instructions for Use" for each device used.
CONTRAINDICATIONS
None known
WARNINGS
Evolution® Shortie RL Controlled-Rotation Dilator Sheath Set should only be used to minimally
enter the vessel.
Do not attempt to negotiate the Evolution® Shortie RL Controlled-Rotation Dilator Sheath Set
past a bend in the vessel as vessel wall or cardiac lead insulation damage may occur.
When using dilator sheaths or sheath sets, do not insert more than one sheath set into a vein at a
time. Severe vessel damage, including venous wall laceration requiring surgical repair, may occur.
When using a Locking Stylet:
Do not abandon a catheter/lead in a patient with a Locking Stylet still in place inside the
catheter/lead. Severe vessel or endocardial wall damage may result from the stiffened catheter/
lead or from fracture or migration of the abandoned stylet wire.
Do not apply weighted traction to an inserted Locking Stylet as myocardial avulsion,
hypotension, or venous wall tearing may result.
Be aware that a lead that has a J-shape retention wire that occupies its inner lumen (rather
than being outside the coil) may not be compatible with the Locking Stylet. Insertion of the
Locking Stylet into such a lead may result in protrusion and possible migration of the J-shape
retention wire.
Weigh the relative risks and benefits of intravascular catheter/lead removal procedures in cases
when:
The item to be removed is of a dangerous shape or configuration,
The likelihood of catheter/lead disintegration resulting in fragment embolism is high or
vegetations are attached directly to the catheter/lead body.
Catheter/lead removal devices should be used only at institutions with thoracic surgical
capabilities.
Catheter/lead removal devices should be used only by physicians knowledgeable in the
techniques and devices for catheter/lead removal.
PRECAUTIONS
When activating the Evolution® Shortie RL Controlled-Rotation Dilator Sheath Set be careful not to cut
or ensnarl lead suture-ties that are attached to the locking stylet targeted lead or to companion leads.
Position the distal tip of the Evolution® Shortie RL Controlled-Rotation Dilator Sheath Set so that it does
not engage companion leads during activation.
Prior to the procedure, consider the size of the catheter/lead in relation to the size of the LEAD
EXTRACTION™ devices to determine possible incompatibility.
If selectively removing catheters/leads with the intent to leave one or more chronic catheters/leads
implanted intact, the non-targeted catheters/leads must be subsequently tested to ensure that they were
not damaged or dislodged during the extraction procedure.
Sheath Set
LR-EVN-SH-9.0-RL
LR-EVN-SH-11.0-RL
2