Error message
17.2.3 Error while charging the product
LED on
LED on
power sup
battery
ply
charger
C-Brace joint unit 17KO1=*
Cause
A current connection to the
component was interrupted
Error
Country-specific plug adapter not fully
engaged on power supply
Non-functional socket
Defective power supply
No connection between battery charger
and power supply
Defective battery charger
Battery is fully charged (or connection
with product is interrupted).
Correction
Check the following points:
•
Distance from the component to the device
•
Charge level of the component's battery
•
Bluetooth of the component switched on? (see
Page 42)
•
Briefly press the
to make the component "visible" for 2 minutes.
•
Component switched on? (see Page 41)
•
If multiple components were stored, was the
correct component selected?
•
Is the component still connected to another
device, and is this connection still active?
Resolution
Check whether the country-specific plug
adapter is fully engaged on the power
supply.
Check
socket
device.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Check whether the charging cable plug is
fully engaged on the battery charger.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Take note of the confirmation signal for
differentiation.
When the battery charger is connected, a
self-test is conducted and confirmed by
beeps/vibration signals.
When these signals are emitted, the bat
tery is fully charged.
If no signals are emitted, the connection
to the product is interrupted.
If the connection to the product is inter
rupted, an authorised Ottobock Service
Centre must inspect the product, battery
charger and power supply.
Appendices
button on the control panel
with
another
electric
55