"C-Brace Setup" adjustment app

8.5 Entering patient data

After each input, a confirmation signal is emitted by the component to confirm the successful transfer of the values.

8.6 Calibration (zero setting)

The individual position of the joint unit has to be calibrated on the patient. If calibration on the patient is omitted,
the swing phase is initiated too early or too late due to incorrect data.
The calibration process must be repeated after every static change on the orthosis (e.g. after adjusting the ankle
joint).
The adjustment process can only be continued after completing the calibration.
INFORMATION
Do not use the hands to help extend the leg.
26
1) Tap the
symbol in the top left corner to open the navigation menu.
2) Tap the "App settings" menu item.
→ The unit settings menu is displayed.
3) Tap the setting in question.
4) Tap the "OK" button to save the setting and close the navigation menu.
Height
Body height of the patient.
Changing the units (see Page 25).
Weight
Body weight of the patient with or without orthosis.
Changing the units (see Page 25)
Knee extension angle
Knee angle of the extension position for which the orthosis was aligned. Slight
hyperextension of the knee joint must be possible starting from this angle.
Frontal Plane angle
Input of the varus/valgus angle for the component. It has to be measured with a
goniometer.
Knee Center to floor
Distance from the knee rotation point to the floor.
Changing the units (see Page 25)
Unilateral
If the orthosis is part of a unilateral fitting, this parameter has to be selected.
Bilateral
If the orthosis is part of a bilateral fitting, this parameter has to be selected.