Sym
Beep
Vibra
bols on
signal
tion sig
the
nal
control
panel
—
—
—
—
3 x long
3 x long Charge level between 5% and 10%
5 x long
5 x long Charge level between 0% and 5%
10 x long 10 x long Charge level 0%
4 x short
at inter
vals of
approx.
65 sec.
(continu
ously)
—
—
—
1 x long
—
—
3 x short Charging mode started (3 sec. after
C-Brace joint unit 17KO1=*
Event
—
Charge level 67% to 100%
Indicates that the battery is fully
charged during the charging pro
cess.
—
Battery is being charged, battery
charge level is less than 34%
—
Battery is being charged, battery
charge level is 34% to 67%
—
Battery is being charged, battery
charge level is 67% to 99%
After the beep and vibration signals,
the product switches to empty battery
mode and then switches off.
—
Charging the battery outside the
allowable temperature range
—
Maintenance required:
E.g. maintenance date exceeded or
defective vibration motor
—
Bluetooth activated
—
Battery charger is connected
connecting battery charger)
Required action
Charge battery soon.
Remaining operating time approx. 1 hour.
Charge battery immediately; the product
will be switched off after the next warning
signal.
Charge the battery.
Check whether the specified ambient con
ditions for charging the battery were met
(see Page 49).
1. In the "C-Brace Setup" adjustment
app (page "Data overview"), check
the total number of steps since the last
maintenance and the last maintenance
date.
2. Check whether a vibration signal is
emitted after connecting the battery
charger.
3. Conduct the self-test again by connect
ing/disconnecting the battery charger.
If the
symbol lights up again and
maintenance is not yet due, the
product must be inspected by an
authorised Ottobock Service Centre.
For disassembly of the joint unit, see
the
section
sembling the joint unit on the orthosis"
(see Page 23)
Appendices
"Assembling/disas
53