Parameter
Sitting function
Volume of acoustic sig­
nal
Pitch of acoustic signal
INFORMATION
Confirmation of the successful transfer of the parameters
A beep and vibration signal is emitted by the orthosis while changing the parameters using the Cockpit app. If the
parameter "Volume of acoustic signal" is set to "0", only a vibration signal is emitted.

10.4.3 Overview of adjustment parameters in MyModes

The parameters in the MyMode "User defined" describe the static behaviour of the orthosis for a specific move­
ment pattern such as cycling. There is no automatically controlled adjustment of the extension and flexion resist­
ance.
The parameters of the MyModes "Training mode" and "Freeze position" are preconfigured and cannot be
changed.
The following parameters can be modified in MyModes:
Parameter
Flexion
resistance
increase
Volume of acoustic sig­
nal
Pitch of acoustic signal
INFORMATION
Confirmation of the successful transfer of the parameters
A beep and vibration signal is emitted by the orthosis while changing the parameters using the Cockpit app. If the
parameter "Volume of acoustic signal" is set to "0", only a vibration signal is emitted.

10.5 Switching the product on/off

In certain cases, e.g. for storage or transportation, the product can be purposely switched off.
CAUTION
Using the product while switched off
Falling due to unexpected product behaviour because of changed damping behaviour.
► Prior to using the product, briefly press the
The button symbol lights up green
C-Brace joint unit 17KO1=*
Adjustment
Cockpit app
range of the
adjustment
O&P profession­
range
al via the adjust­
ment app
0/Off - deactiv­
ated
1/On - activated
0 to 100
0 to 100
1500 Hz
1500 Hz
3000 Hz
3000 Hz
Adjustment app
Cockpit app
setting range
adjustment
range
0 to 100
+/- 10 of the con­
figured value
0 to 100
0 to 100
1500 Hz
1500 Hz
3000 Hz
3000 Hz
button on the control panel to check whether it is switched on.
when the product is switched on.
Meaning
Activation/deactivation of the sitting function.
This function needs to be enabled in the
adjustment app.
Volume of beep signal for confirmation tones
(e.g. when switching MyModes). The "0" set­
ting deactivates the audible feedback signals.
However, warning signals are still generated
if errors occur.
Pitch of beep signal for confirmation tones
Meaning
Value for speed at which flexion resistance
increases as the knee angle increases
This parameter can only be changed in the
MyMode "User defined".
Volume of beep signal for confirmation tones
(e.g. when switching MyModes). The "0" set­
ting deactivates the audible feedback signals.
However, warning signals are still generated
if errors occur.
Pitch of beep signal for confirmation tones
Use