"C-Brace Setup" adjustment app
8.11 Configuration of the Cockpit app
30
Parameter "Volume of acoustic signal"
Setting for the volume of the acoustic signal generator for the confirmation sig
nals. Warning signals that indicate an error in the system always have the
highest volume.
Parameter "Pitch of acoustic signal"
Setting for the pitch of the acoustic signal for the confirmation signals.
Parameter "Safety mode flexion resistance"
When a critical error occurs or if the battery is drained, the component
switches to safety mode and swing phase initiation is no longer possible. Fur
ther information about safety mode (see Page 46).
The flexion resistance is set to constant with the value of the parameter "Safety
mode flexion resistance" in the following cases:
•
The component is in safety mode
•
The component has been switched off
•
The component battery empty
•
The component is being charged (battery charger is connected to the com
ponent)
Button "Show Bluetooth PIN"
A Bluetooth PIN is required to establish a connection with the Cockpit app.
This PIN is provided on a card included in the scope of delivery. If this card is
no longer available, the Bluetooth PIN can be displayed by tapping this button.
The Bluetooth PIN is only valid for the currently connected component.
Menu language of Cockpit app
Select the user interface language of the Cockpit app. Only the languages that
can be selected here are available for the Cockpit app. Subsequently changing
the language via the mobile device is not possible.
Name of the component
Enter an individual name of the component for identification in the Cockpit app.
Button "Transfer configuration of Cockpit app"
The changes (Menu language of Cockpit app, Name of the component)
are not saved in the component until this button is tapped.
The patient has to sit or stand with support during the data transfer.
C-Brace joint unit 17KO1=*