Flow Connections

Tools required
Description
Parts from the HPLC Tool-Kit
Parts required
Description
Parts from the Accessory kit
Preparations
• Sampler is installed in the LC system
Toxic, flammable and hazardous solvents, samples and reagents
WA R N I N G
The handling of solvents, samples and reagents can hold health and safety risks.
1 Connect the pump outlet capillary to port 1 of the injection valve.
2 Connect column-compartment inlet capillary to port 6 of the injection
3 Connect the corrugated waste tube to the solvent waste from the leak plane.
4 Ensure that the waste tube is positioned inside the leak channel.
Do not extend the waste capillary of the autosampler. The siphoning effect might empty the
N O T E
complete seat capillary introducing air into the system.
Agilent 1260 Infinity Standard Autosampler User Manual
When working with these substances observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in
the material handling and safety data sheet supplied by the vendor, and follow good
laboratory practice.
The volume of substances should be reduced to the minimum required for the
analysis.
Do not operate the instrument in an explosive atmosphere.
valve.
Installing the Autosampler

Flow Connections

3
35