8 Program values
8.1 Dose information
Program
Maximum effective
exposure times
P1
14.2s
P1L
8.0s
P1R
8.0s
P1A
14.2s
P1C
14.2s
P20
11.3s
BW10
8.8s
TM1.1+TM1.2
6.4+6.4s
68
8.1

Dose information

The dose area product for the parameter combinations proposed by
Preface
Sirona has been calculated already. The DAP value can be used without
any further calculations.
8.1.1

Dose area product parameters for Panorama images

The dose area product (DAP) values were measured in a CT ionization
chamber.
Dose order code number 4A (8 mA/16 mA series)
DAP ORTHOPHOS XG 3
Specification of the dose-area product (DAP/absorbed dose) for
panoramic and temporomandibular joint views:
Factory-programmed values
DAP
2
mGycm
kV/mA
62/8
64
62/8
36
62/8
36
62/8
64
62/8
64
62/8
51
62/8
40
68/8
69
DAP ORTHOPHOS XG 3
Operating Instructions ORTHOPHOS XG 3
DAP
2
mGycm
kV/mA
kV/mA
64/8
68
69/15
64/8
39
69/15
64/8
39
69/15
64/8
68
69/15
64/8
68
69/15
64/8
54
69/15
64/8
42
69/15
71/8
75
73/15
Sirona Dental Systems GmbH
DAP
DAP
2
mGycm
mGycm
kV/mA
147
73/15
164
83
73/15
92
83
73/15
92
147
73/15
164
147
73/15
164
117
73/15
130
91
73/15
101
147
77/14
151
60 50 681 D3352