Replacement
Replacing the Fuses:
Rating of Fuses
Thermo Fisher Scientific
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Work
e two power fuses are located above the main switch [a] on the rear
side of the microtome.
To replace the two fuses, turn o the main switch [a] of the instru-
ment and unplug it.
en open the cover [b] using a at screwdriver.
Pull out the fuse holders and replace the fuses with new ones.
Put back the fuse holder completly and close the cover [b].
For power requirements 100 – 240 V, 2 fuses T2,5AH slow-blow
Maintenance and Care

Replacement Work

[b]
[a]
Operation Manual HM335S
67