Thermo Scientific Microm HM355S Operation Manual page 60

Rotary microtome
Hide thumbs Also See for Microm HM355S:
Table of Contents

Advertisement

Available languages

Available languages

Installing the Large Field
Magnifier on the Microtone
Thermo Fisher Scientific
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Connect the integrated light to the power outlet with the power cord [b]
and turn it and o with the switch [c].
C
!
AUTION
Before using the large eld magni er for the rst time, please check if
the voltage conditions at the installation site comply with the power
requirements and frequency noted on the supply unit of the magni er.
Please use the enclosed fastening elements to attach the magni er to the
microtome.
Remove the cover plate on the upper side of the microtome hood.
Remove the front plastic caps from the upper side of the housing.
Carefully introduce the four sleeves [g] into the drilled holes with
guidance of the screws [e].
C
!
AUTION
Please be sure that the sleeves do not fall into the interior of the micro-
tome!
Note!
Make sure the plate [f] is mounted in such a way that the peg [d] is closer to
the front of the microtome!
Put the plate [f ] on the sleeves and fasten it to the microtome
with four screws [e].
Put the movable plastic socket, which is located at the lower side of
the magni er, on the peg [d].
e magni er can be adjusted forwards and backwards for the most
comfortable viewing position.
Note!
If the large eld magni er is not used, slightly raise it and turn it sideways
on the peg.
Operating Instructions
Large Field Magnifier
Operation Manual HM335S
59

Hide quick links:

Advertisement

Chapters

Table of Contents
loading

Table of Contents