replacement
replacing the Fuses:
rating of Fuses
Thermo Fisher Scientific
Work
The two power fuses are located above the main switch [a] on the rear
side of the microtome.
•
To replace the two fuses, turn off the main switch [a] of the instru-
ment and unplug it.
•
Then open the cover [b] using a flat screwdriver.
•
Pull out the fuse holders and replace the fuses with new ones.
•
Put back the fuse holder completly and close the cover [b].
For power requirements 100 – 240 V, 2 fuses T2,5AH slow-blow
Maintenance and Care
Replacement Work
[b]
[a]
Operation Manual HM335S
67