Working with the Microtome
Section Instructions
Operation Manual HM335S
62
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Between these two limit cases, the correct knife angle must be deter-
mined. A clearance angle between facet angle and block surface of
approx. 5° is normal.
C
!
AUTION
Di erent microtome producer use di erent scaling for their clearance
angle adjustment, resulting from di erent calculation bases.
ance angle might refer to the facet surface or to the angle between knife
main surface and block surface. As the facet has a divergence from the
knife main surface of approx. 5°, the resulting di erence is approx. 5°.
T
,
10°.