6.1.2

Uncritical surfaces including monitor

NOTE
Do not allow liquids to penetrate into the ventilation slots!
NOTE
Never use corrosive cleaning agents, wax or solvents.
1. Use a new absorbent cotton gauze, which is moistened with one of
the cleaning agents listed in the "Wiping Disinfectants  [→ 62]"
section, or a moistened cloth to disinfect the non-critical contact
surfaces.
2. To clean and disinfect the monitor, follow the following steps, so
that the monitor is frozen and you are not able to accidentally
activate a function on the screen during wiping:
a) Press simultaneously on the left and right button on the touchpad
for approx. 1 second. The screen turns dark and "device locked"
appears.
b) Wipe the monitor.
c) Unlock the monitor with the foot control.
3. Then remove the absorbent cotton gauze or wiping cloth.
Please ensure that no colored cloths are used for this as these can
result in discoloration of the surfaces, e.g. when used in combination
with disinfectants.
Protection against medicaments
Due to their high concentrations and the substances they contain, many
medicaments can dissolve, etch, bleach or discolor surfaces.
NOTE
The only way to prevent damage is to wipe off medicaments
immediately with a damp cloth and a cleaning agent!
6 Maintenance
6.1 Cleaning, disinfection and sterilization
63