Maintenance and inspection

Hygiene

9.1
Disinfect and sterilize accessories prior to initial use and after each
patient.
9.1.1

Care, cleaning agents, and disinfectants

NOTE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning, and disinfecting agents approved by
Dentsply Sirona!
A continuously updated list of approved agents can be downloaded
from the Internet on the online portal for technical documents. You can
reach this portal at the address:
www.dentsplysirona.com/manuals
.Click on the menu item "General documents" and then open the "Care,
cleaning and disinfection agents"  document.
If you do not have access to the internet, please contact your dental
depot to order the list (REF 59 70 905).
9.1.2

USB interface and sensors

NOTE
Liquids can get into the USB module or the sensor during cleaning
and disinfection. The plug contacts may become wet.
The USB module, sensor and PC can be damaged or destroyed by a
short circuit.
➢ Before cleaning and disinfection, remove the USB plug on the PC
or USB hub. The system must be disconnected from the mains.
Also remove the sensor plug from the USB module.
➢ The USB module and the sensor must not be thermal disinfected,
sterilized or immersed in disinfectant. They must not be
disinfected or sterilized with radiation. Apply wipe disinfectant
only.
➢ Never spray plug connections with disinfectants or cleaning
solutions. Make sure that the plug contacts do not become wet.
NOTE
Drugs have a chemical reaction with the surface of the unit.
Due to their high concentrations and the substances they contain,
many drugs can dissolve, etch, bleach, or stain surfaces.
➢ Wipe any drug residues off the unit immediately with a moist,
white cloth!
ü All plug connections are disconnected.
1. Clean the sensor, the sensor cable and the USB module with a
cloth soaked in soapy water. After this, dry the components with a
lint-free cloth.