:MEASure:TMAX

Command Syntax
:MEASure:TMAX [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:TMAX command installs a screen measurement and starts
an X- at- Max- Y measurement on the selected waveform. If the optional
source is specified, the current source is modified.
The :MEASure:TMAX command is obsolete and is provided for backward compatibility to
N O T E
previous oscilloscopes. Use the :MEASure:XMAX command (see
Query Syntax
:MEASure:TMAX? [<source>]
The :MEASure:TMAX? query returns the horizontal axis value at which the
maximum vertical value occurs on the current source. If the optional
source is specified, the current source is modified. If all channels are off,
the query returns 9.9E+37.
Return Format
<value><NL>
<value> ::= time at maximum in NR3 format
See Also
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
586)
"Introduction to :MEASure Commands"
":MEASure:TMIN"
on page 528
":MEASure:XMAX"
on page 295
":MEASure:XMIN"
on page 296
Obsolete and Discontinued Commands
page
295) instead.