5
Commands by Subsystem
Table 31 Service Request Enable Register (SRE) (continued)
Bit
5
4
3
2
1
0
Query Syntax
*SRE?
The *SRE? query returns the current value of the Service Request Enable
Register.
Return Format
<mask><NL>
<mask> ::= sum of all bits that are set, 0,..,255;
See Also
110
Name
Description
ESB
Event Status Bit
MAV
Message Available
---
---
MSG
Message
USR
User Event
TRG
Trigger
an integer in NR1 format
"Introduction to Common (*) Commands"
"*STB (Read Status Byte)"
"*CLS (Clear Status)"
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
When Set (1 = High = True), Enables:
Interrupts when enabled conditions in the
Standard Event Status Register (ESR) occur.
Interrupts when messages are in the Output
Queue.
(Not used.)
Interrupts when an advisory has been displayed
on the oscilloscope.
Interrupts when enabled user event conditions
occur.
Interrupts when a trigger occurs.