:TRIGger:LIN:SYNCbreak

Command Syntax
:TRIGger:LIN:SYNCbreak <value>
<value> ::= integer = {11 | 12 | 13}
The :TRIGger:LIN:SYNCbreak command sets the length of the LIN sync
break to be greater than or equal to 11,12, or 13 clock lengths. The sync
break is the idle period in the bus activity at the beginning of each packet
that distinguishes one information packet from the previous one.
Query Syntax
:TRIGger:LIN:SYNCbreak?
The :TRIGger:LIN:STANdard? query returns the current LIN sync break
setting.
Return Format
<value><NL>
<value> ::= {11 | 12 | 13}
See Also
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
586)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 360
":TRIGger:LIN:SIGNal:DEFinition"
":TRIGger:LIN:SOURce"
on page 354
on page 542
on page 409