:TRIGger:UART:BITorder

Command Syntax
:TRIGger:UART:BITorder <bitorder>
<bitorder> ::= {LSBFirst | MSBFirst}
The :TRIGger:UART:BITorder command specifies the order of transmission
used by the physical Tx and Rx input signals for the serial decoder and/or
trigger when in UART mode. LSBFirst sets the least significant bit of each
message "byte" as transmitted first. MSBFirst sets the most significant bit
as transmitted first.
Query Syntax
:TRIGger:UART:BITorder?
The :TRIGger:UART:BITorder? query returns the current UART bit order
setting.
Return Format
<bitorder><NL>
<bitorder> ::= {LSBF | MSBF}
See Also
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
586)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 360
":TRIGger:UART:TYPE"
":TRIGger:UART:SOURce:RX"
":TRIGger:UART:SOURce:TX"
on page 354
on page 440
on page 438
on page 439