Transfer Experiments, Templates, And Results Data - Applied Biosystems QuantStudio 6 Flex User Manual

Table of Contents

Advertisement

Transfer
experiments,
templates, and
results data
QuantStudio
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide
6. Modify the experiment parameters as needed.
• Add a stage or step to the thermal profile, tap the stage or step to the left of
where you want to add the stage or step, then tap
• Add a melt curve to the end of the thermal profile, tap
• Change the time or temperature of a stage or step, touch the
time/temperature field of the stage or step, modify the settings as desired,
then tap Close.
• Change the cycle parameter of a stage, tap the cycle field, modify the setting
as desired, then tap Close.
• Delete a stage or step from the thermal profile, tap the stage or step you
want to remove, then tap
7. When finished modifying the parameters, tap
8. In the Save Experiment screen, tap each field to set the experiment, name,
reaction volume, barcode, and any additional information to save to the
experiment
9. When finished, tap
10. In the Start Run screen, tap each field as needed to modify the associated
parameter, then tap
Note: When the run is complete, tap
the experiment results to a computer if the QuantStudio
PCR System is connected to a network, or you can copy the data to a USB drive
(see "Transfer experiments, templates, and results data" on page 85).
You can transfer experiments, templates,
and results data to/from the
QuantStudio
6 and 7 Flex Systems using
a USB flash drive. Before transferring
data, you must plug the drive into the
USB port below the touchscreen.
IMPORTANT! Do not use the USB ports
on the rear panel of the QuantStudio
and 7 Flex Systems. The rear USB ports
are only for use by Thermo Fisher
Scientific personnel to service the
instrument.
Transfer templates from a USB drive
1. Plug a USB drive into the USB port below the touchscreen.
2. If the instrument is in standby, touch the QuantStudio
touchscreen to activate it, then press
3. In the Main Menu, touch
Appendix A Manual instrument operation
Operate the instrument from the touchscreen
Delete.
Save & Start Run to start the experiment.