Chapter 2 Data Sheet: Calibration Verification Test

Data Sheet: Calibration Verification Test

18
Thermal Cycler Temperature Verification System For GeneAmp
When you run the calibration verification test, record the setpoint values for well A6
on this data sheet. At the end of the test, compare the values that are displayed on the
thermal cycler with the values that are recorded here.
You can photocopy this page.
Note:
Date
Tested by
Thermometer
Probe serial no.
serial no.
Setpoint value: Well A6
85 °C
®
PCR Systems 2720, 9600, 9700, 9800,
®
and Veriti
System User Guide
45 °C