Test Results

Perform the following steps to calculate the results of the test.
You will need to refer to the calibration label that you pasted into
Figure 8-4 below, for the High and Low Offset values.
Figure 8-4 GeneAmp PCR System 9600 Calibration Label.
Note
laboratory, make sure that the serial number on the calibration label
matches the serial number on the instrument you are testing.
1
8-10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
PASTE THE CALIBRATION LABEL HERE.
If you have more than one GeneAmp PCR System 9600 in your
Use the following formula to calculate the average block
temperature at the 95-degree hold:
Block Average at 95°C = T(95) – High Offset