We recommend several protective devices to protect the instrument in environments
with large voltage and power fluctuations.
Power line regulator
We recommend the use of a 1.5-kVA power line regulator in areas where the supplied
power fluctuates in excess of ± 10% of the normal voltage. Power fluctuations can
adversely affect the function of the instrument.
Note: A power line regulator monitors the input current and adjusts the power
supplied to the instrument or computer. It does not protect against a power surge or
failure.
Uninterruptible power supply (UPS)
We recommend the use of a 1.5-kVA uninterruptible power supply (UPS), especially
in areas prone to power failure. Power failures and other events that abruptly
terminate the function of the QuantStudio
possibly damage the computer or the instrument.
IMPORTANT! UPSs provide power for a limited time. They are meant to delay the
effects of a power outage, not to serve as replacement power sources. In the event of a
power loss, power off the instrument and the computer, unless you expect to regain
power within the battery life of the UPS.
Surge protector
We recommend the use of a 10-kVA surge protector (line conditioner) in areas with
frequent electrical storms or near devices that are electrically noisy, such as
refrigerators, air conditioners, or centrifuges. Short-duration, high-voltage power
fluctuations can abruptly terminate the function of, and thereby damage the
components of, the computer and the instrument.
Note: A dedicated line and ground between the instrument/computer and the
building's main electrical service can also prevent problems caused by power
fluctuations.
The QuantStudio
6 and 7 Flex Systems include a suite of software applications that
can be used to calibrate, run, automate, and integrate the instrument into a laboratory
workflow. The installation of the software contains the components described in the
following sections; however, additional software may be available for the
QuantStudio
6 and 7 Flex Systems. Go to thermofisher.com/qpcrsoftware for a
complete list of compatible software, and for updates and patches for the system
software and firmware.
The requirements for the computer used to operate the QuantStudio
Systems can vary depending on the version of the QuantStudio
Software that you are running. To determine the computer requirements for your
system, check the software release notes at thermofisher.com/qpcrsoftware. The
release notes are on the same page as the link for the software download.