Appendix B Maintenance; Decontaminate The Sample Block - Applied Biosystems QuantStudio 6 Flex User Manual

Table of Contents

Advertisement

B

Decontaminate the sample block

QuantStudio
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide
Decontaminate the sample block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Replace the halogen lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Replace the instrument fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Software licenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Update the QuantStudio
Change the sample block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Change the heated cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Change the plate adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
IMPORTANT! This chapter contains all user service procedures for the QuantStudio
6 and 7 Flex Systems. Procedures other than those described in this document must be
performed by a qualified Thermo Fisher Scientific Service Engineer.
Perform this procedure to eliminate fluorescent contaminants from the sample block.
Contamination is generally evident in failed background calibrations where one or
more wells consistently exhibit abnormally high signals.
CAUTION! PHYSICAL INJURY HAZARD. Do not remove the QuantStudio
6 or 7 Flex Real-Time PCR System cover. There are no components inside the
instrument that you can safely service yourself. If you suspect a problem,
contact a Thermo Fisher Scientific Service Representative.
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the
sample block can reach 100°C. Allow the sample block to cool to room
temperature before performing the procedure.
CAUTION! Before using a cleaning or decontamination method other than
those recommended by Thermo Fisher Scientific, confirm with Thermo Fisher
Scientific that the proposed method will not damage the equipment.
Real-Time PCR Software . . . . . . . . . . . . . . . . . . . . . . 104
Real-Time PCR Software . . . . . . . . . . . . . . . . . . . . . 106
operating system . . . . . . . . . . . . . . . . . . . . 107
Maintenance
97