Service Menu; Logs Menu - Covidien Nellcor GR101704 Service Manual

Bedside respiratory patient monitoring system
Table of Contents

Advertisement

a.
b.
Restore Factory Defaults
Proceed only if institutional defaults require replacement with the standard
factory defaults.
To restore institutional defaults back to factory defaults
Access the service function menu. Reference
1.
Select SETTINGS.
2.
Select RESTORE FACTORY DEFAULTS.
3.
Select the desired option.
4.
a.
b.
Select the desired option.
5.
a.
b.

Service Menu

6.5.2
Reference
ibration instructions. Reference
for software or firmware upgrade instructions.

Logs Menu

6.5.3
Error, event, and software update logs remain in flash until the monitoring
system overwrites them and will survive both power cycles and software
upgrades. If a user chooses to "clear" a log, the monitoring system renames
the current log files, rather than deleting them. Possible scenarios for either file
corruption or a partial loss of data in a log would include backdating the mon-
itoring system when it has already recorded data to an existing date or a power
interrupt during retrieval of a log.
Service Manual
Press SAVE CHANGES to retain the change.
Press CANCEL to leave the default as it was.
Press YES to restore the factory defaults.
Press CANCEL to retain the factory defaults.
Press BACK to return to the previous submenu.
Press EXIT MENU.
Monitoring Screen
To access service function
Calibration, p. 10-44, for monitoring screen cal-
Software and Firmware
menus, p. 6-18.
Upgrades, p. 10-45,
Service Mode
6-31