Respiration Rate Safety Information

1.4.3
Upon connection of the respiratory sensor to the monitoring system, you are
prompted to review on-screen Respiration Rate safety information.
To review the safety information
Press EXIT, then press HELP.
1.
Select RR Safety Information from the Help options, and review the safety infor-
2.
mation.
Press Exit.
3.

Menu Options

1.4.4
Most menu options remain the same as those listed in the Operator's Manual.
Menu options specific to Respiration Rate are described below.
ALARM LIMITS Menu - Respiration Rate Alarm Limits
The ALARM LIMITS menu includes an option for setting upper and lower res-
piration rate alarm limits. An alarm is indicated each time a patient's respiration
rate violates either limit.
By default, the upper alarm limit is 28 breaths per minute, and the lower alarm
limit is 8 breaths per minute. If changing these values, be sure to observe all
warnings and cautions presented in the Operator's Manual regarding setting
appropriate alarm limits.
Note:
The ability for users to adjust alarm limit settings can be enabled or disabled by
qualified service personnel as described in the Service Manual.