GE HEALTHCARE
REV 9
Step
From previous page
Manual (SM)
D Ultranet SA Collimator + RAFL−S
D C40E
A34
D SG60/SG100
1 Person
DD TOMO−LINK CALIBRATION
A35
D Perform C1, C2, C3, C4, C5 ,C7
Calibration Procedures
D Drive Power amp adjustment
1 Person
System Service Manual − SYSTEM COMPAX 40E, CE Marked
Description
Ultranet
Collimator
Auto/Manual
Auto (SA)
CALIBRATION
CALIBRATION
Continues on next page
with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)
Í Í
Í Í
Supporting Documents
Done
Í Í
2 Hours
Í Í
Í Í
Í Í
1 Hour
28
DIRECTION 2154260−100
Refer to Inspection Report 2179651 TDR,
section 7.3, or 2167647 TDR or 2172166 TDR,
section 6.3
Proceed with required calibrations labelled:
See ULTRANET SA + RAFL−S SM
Dir. Y00W20L.04, Chaps 3, 4 and 6.
(Use extension cable/support provided to
connect RAFL keyboard and fix it on front
of cabinet − Must be removed after the
CALIBRATION is completed).
See Job Card IST005 to open the
RAFL−S rack
Refer to Inspection Report 2179651 TDR
section 5.2
See SM Tomo−Link Dir. 2133821−100,
Section 16 and 20.
.