Job Card Ist 001 - GE Compax 40E System Service Manual

Table of Contents

Advertisement

GE HEALTHCARE
REV 9
SYSTEM COMPAX 40E
X−RAY TUBE & COLLIMATOR
Purpose:
INSTALLATION ON XT SUSPENSION
Time:
SECTION 1
SPARES REQUIRED
SECTION 2
EQUIPMENT REQUIRED
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PRELIMINARY STEPS
SECTION 5
INSTALLATION PROCEDURES
5.1
Configuration During HLA
For XT Suspension installation in Europe the Collimator lamp voltage is configured by connecting
an electrical jumper to J10 (24V
a 230VAC input line is on XT transformer (refer to System Inspection Report and Illustration 1).
Note:
An XT Suspension integrated in the US is furnished with jumper connected to J9.
System Service Manual − SYSTEM COMPAX 40E, CE Marked
with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)
WITCH THE POWER OFF ON XT SUSPENSION BEFORE CHANGING TAP
SELECTION ON XT CONTROL BOARD.
) for the Philips lamp or J11 (26V
AC
63
DIRECTION 2154260−100

Job Card IST 001

Version No.: 2
Date: August 7, 1997
Personnel:
1
for the GE type lamp, when
AC)
1 of 4