GE HEALTHCARE
REV 9
ILLUSTRATION 7−5
GROUNDING INTERCONNECTING MAP (SYSTEM MPH 50/65/80 GENERATOR AND RS85 TUBE STAND WITH COMPAX 40E−MANUAL AND SG100)
Note:
All grounding cables are yellow−green.
MPH 50/65/80 CABINET
MAIN GROUND
TUBESTA
MPH A1 A2
POWER
Stator Connector
SUPPLY
Panel
2138905 ( Power Supply Cable )
27235
1.1
OPTIONAL
REFER TO JOBCARD IST002
IN APPROPRIATE VERSION
MPH SERVICE MANUALS
FOR GROUND PLUG LAYOUT
MPH V1 SM:
2137354−100
MPH V2/V3 SM: 2165156−100
2105836
Printer Ground Cable
(Supplied with Printer)
LEGENDE
S
SIGNAL POINT GND
CHASSIS
PROTECTIVE
EARTH TERMINAL
REFERENCE TERMINAL
EQUIPOTENTIAL TERMINAL
MIS RUN − FIELD
000
SUPPLIED WIRE
SAFETY GROUND WIRE
A
SIGNAL GROUND WIRE
B
SAFETY & SIGNAL
C
GROUND WIRE
MIS RUN − CABLE
000
REFER TO OPTION
MAPS FOR MIS NUMBER
FACTORY INSTALLED WIRE
MPH A6 A8
Plugs Panel
GND
POWER SUPPLY
MPH A6 A6
MPH A6 A4
II Sensor Board
Wall Bucky
RS 85
24VDC
ND
MPH A6 A3
MPH A6 A5
Distribution
Table Tomo
Board
Collimator
MPH A4
Front Electronic Panel
MPH A3 A2
7.1
Main Power
MPH A3
MPH A5
A1
Real Electronic Panel
DC Power
MPH A3 A3
MPH A3
Inverter Board
FL1
MPH
Ground Plug
A1
I +
I −
MPH A2
HV TANK
2165548
POWER SUPPLY CABLE
2.1
6.1
45553701
(Console Ground Cable)
MPH A8
MPH A7
Program−X
Control Console
GND STUD
GND STUD
MPH A9
Printer−X
109
2165548
POWER SUPPLY CABLE
7.1
27238
2140362 ( SG Ground Cable )
2140362
27302
27301
4.1
RS85 Ground Cable
2140362 ( TABLE Ground Cable )
3.1
27300
System Service Manual − SYSTEM COMPAX 40E, CE Marked
with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)
2183955
GND
RS85 TOWER
5.1
8.1
27305
X−RAY TUBE
GND X−RAY TUBE
+
SA/SM Ultranet
COLLIMATOR
RS85 TUBE STAND
COMPAX 40E TABLE
TB1−GND
DIRECTION 2154260−100
I/O PANEL GND
SG 100 WALL STAND
Need help?