GE HEALTHCARE
REV 9
X−RAY TUBE & COLLIMATOR
INSTALLATION ON XT SUSPENSION
5.2
Collimator Lamp voltage Selection Procedure
Before starting this procedure ensure that the XT control board and the Collimator lamp are
accessible (refer to Collimator SM for instructions).
1. Select on digital multimeter:
− V
AC
− Auto range (or manual range > 240 V
2. Measure the actual voltage at Collimator lamp bulb base with the lamp switched ON.
3. With The GE type lamp, ensure that voltage is greater than 23V when the electrical jumper
J11 is in place.
4. With The Philips type lamp, if the voltage is within 21.5−V
of the electrical jumper position next to J9 or J10 cannot be used to set the voltage closer to
22.5V
AC
your system.
If so, then move jumper from J10 to J9. (Refer to table below for tap selection voltages).
Terminal blade
Selected in HLA
5. If the voltage is out of the range 21.5 − 24 V
(between J100−1 & J100−5). If it is not 230 V
Dir. 2137354−100 and correct line voltage setting on generator.
If the input line is correct, move the electrical jumper to a more appropriate position in order to
keep the voltage within 21.5 and 24 V
greater than 23 V
6. Record actual final voltage reading at Collimator lamp bulb base only if you have to do HHS
measurement on your system.
System Service Manual − SYSTEM COMPAX 40E, CE Marked
function
(always above 21. V
), this depends on the actual room line voltage available on
AC
Terminal blade
straight
straight
J7
J8
J7
J9
J7
J10
J7
J11
for the GE lamp.
AC
Actual Voltage
with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)
Job Card IST 001
).
AC
AC
Field Lamp TAP select
20V
AC
22V
AC
24V
AC
26V
AC
check the input line on the XT control board
AC,
+/− 12.5%, redo IST 02 of SM MPH 50/65/80
AC
and idealy close to 22.5 V
AC
VAC
65
DIRECTION 2154260−100
24V
range be sure that one
AC
TAP
TAP
TAP
TAP
for the Philiips lamp and
AC
3 of 4