GE HEALTHCARE
REV 9
MPH 50/65/80 CABINET
CABLE #
10
MPH A6 A4 MODULE
(Wall Bucky Board)
CAUTION
11
MPH A1 MODULE
(AC Distribution Panel)
12
MPH A6 A5 MODULE
(Table Tomo Board)
CAUTION
13
MPH A6 A3 MODULE
(Distribution Board)
XT SUSPENSION
14
Control
Board 3A1
connector support
System Service Manual − SYSTEM COMPAX 40E, CE Marked
O Vac
XJ3−7
XJ3−6
H
Bucky Start
XJ3−4
115/230Vac
XJ3−9
XJ3−2
RTN
Exp Order
XJ3−1
shield
clamp
VERIFY THE BUCKY SUPPLY 115/240 SETTING
230Vac
:
F18−2
O Vac
F19−2
GND
Ground Plug
O Vac
XJ10−7
XJ10−6
H
Bucky Start
XJ10−4
115/230Vac
XJ10−9
C40E TABLE BUCKY CABLE
XJ10−2
RTN
Exp Order
XJ10−1
shield
clamp
VERIFY THE BUCKY SUPPLY 115/240 SETTING
Right area
XJ15−F
IC output
XJ15−D
+15VDC
XJ15−A
Center area
H
XJ15−K
−15VDC.
XJ15−C
Left area
XJ15−J
C40E Table Ion Chamber CABLE
Start Int.
XJ15−H
Relay RTN
XJ15−E
XJ15−B
0VDC
XJ15−P
+ HV
shield
clamped
−10V ref.
J101−9
SID pot.
J101−10
SID pot. rtn
J101−8
:
Ref, rtn coll 1
J101−7
Ver. Detent
J101−3
Ver. Move
J101−35
J101−16
XT ok for Tomo
J101−12
Ordo Lock Release.
J101−11
P21V
shield
with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)
MIS 27221
:
2145641
SG60/100 BUCKY CABLE
MIS 27227
2167144
TABLE Power Supply CABLE
MIS 27228
:
2145602
MIS 27229
:
2145651
MANUAL
MIS 10705
:
Ref, rtn coll 1
46−329341G2
SID CABLE
XT ok for Tomo
Ordo Lock Release.
51
DIRECTION 2154260−100
SG60/100 VERTICAL
BUCKY STAND
O Vac
J42−X
Bucky Start
J42−N
220Vac
J42−D
I / O PANEL
RTN
J42−d
Exp Order
J42−b
shield
connector plate
COMPAX 40E TABLE
230Vac
TB1 (Phase)
O Vac
TB1 (Neutral)
I / O PANEL
GND
TB1 (Ground)
O Vac
J42−X
Bucky Start
J42−N
220Vac
J42−D
I / O PANEL
RTN
J42−d
Exp Order
J42−b
shield
connector plate
Right area
J40−A
IC output
J40−B
+15VDC
J40−C
Center area
J40−D
−15VDC.
J40−F
Left area
I / O PANEL
J40−H
Start Int.
J40−J
Relay RTN
J40−E
0VDC
J40−L
+ HV
J40−R
shield
isolated
−10V ref.
J43−9
SID pot.
J43−10
SID pot. rtn
J43−8
J43−7
Ver. Detent
J43−3
Ver. Move
I / O PANEL
J43−35
J43−16
J43−12
P21V
J43−11
shield
connector support
Need help?