GE HEALTHCARE
REV 9
SECTION 1
INTRODUCTION
1-1
Safety
PREVENTION OF IONIZING RADIATION
−
It is mandatory to wear a film badge.
−
Remain as little time as possible in or near of an X−ray field.
−
When working near X−rays, it is mandatory to wear a lead (Pb) apron.
−
The kV and mA setting values must be as low as possible.
−
The X−ray field setting must be as low as possible (a beam limiter must be used).
−
The operator must stand as far away as possible from the X−ray beam.
PREVENTION OF ELECTRICAL HAZARDS
−
Do not open electrical cabinets if power supply is on.
−
Do not disconnect a ground wire.
−
Before disconnecting HV cables, discharge them with a voltage divider.
−
Be careful of capacitors which may be charged (specially those HV units).
PREVENTION OF ACCIDENTS CAUSED BY MOVING PARTS
When adjusting mechanical sub−assemblies, be careful of:
−
movements starting up.
−
transmission parts (chain, gears, ....).
EQUIPMENT PROTECTION
To avoid accidental damage to the equipment, it is mandatory to cut the electric power supply before:
−
connecting or disconnecting a printed board or a connector.
−
connecting or disconnecting a test probe.
System Service Manual − SYSTEM COMPAX 40E, CE Marked
with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)
CHAPTER 1 − PRELIMINARIES
ALL SERVICE OR SPARES REPLACEMENT ON THIS SYSTEM MUST BE
CARRIED OUT BY QUALIFIED TECHNICIANS ONLY. FOLLOW THE
DIRECTIONS IN THIS MANUAL CAREFULLY.
1
DIRECTION 2154260−100