SignalShark User Manual
The data set is saved to the current working folder.
For more information about setting a working folder see
Changing the general display and tone settings
(Display and Tone)
Here you can change general display and tone settings such as the zero point of the Y-axis,
the range of the Y-axis or the axis unit (the available settings depend on the current view).
Lmax
Here you can do the following:
• Set the maximum power level of the graph
• Set the Ref.L. -> Lmax function On or Off
• Select Default Range
• Select the Unit
To set the maximum power level:
1. Tap the Lmax button.
2. Enter the required value and tap the unit button.
The setting is accepted immediately and the input window is closed.
3. Press the Esc key to close the menu.
To select the Unit:
1. Tap the Unit button.
The Unit dropdown menu opens.
2. Select one of the units (dBm, dBV, dBmV, dBuV) and close the menu by tapping the x
button.
Lrange
Here you can do the following:
• Set the range of the y-axis
• Select the default range
• Select a unit
To set the y-axis range:
1. Tap the Lrange button.
2. Enter the required value and tap the unit button.
The setting is accepted immediately and the input window is closed.
To select the unit:
1. Tap the Unit button.
The Unit dropdown menu opens.
2. Select one of the units (dBm, dBV, dBmV, dBuV) and close the menu by tapping the x
button.
172