Enabling/disabling oversampling (CBW only)
You selected the desired view and selected CBW as filter type.
1. Tap the CBW button.
The context menu opens.
2. Tap the Oversampling button to switch it On or Off.
3. Press the Esc key to close the context menu.

Setting pre average (Pre Avg.)

Setting pre average / switching it off
1. Tap the Pre Avg. button.
2. Enter the required value and tap the unit button.
The setting is accepted immediately and the input window is closed.
3. Set the value to zero to switch pre average off.
Assigning channel bandwidth to pre average (CBW –> Pre Avg.):
Tap the CBW –> Pre Avg. button to switch it on.

Setting the measurement time (Meas. Time)

Here you can set the measurement time. The measurement or detector time defines the time
range used by the internal detectors to calculate their values (+Pk, RMS, -Pk, ...).
To set the measurement time:
You selected the desired view.
1. Tap the Meas. Time button.
2. Enter the required value and select a unit.

Configuring the measurement performance (Config)

Here you can define how measurements are performed.
1. Tap the Config button.
2. Then you can do the following:
Setting the magnentic declination (Magn. Decl.)
1. Tap the button and enter the required value.
2. Tap the OK button.
3. Press the Esc key or tap the header to close the menu.
Defining the stop behavior (Stop Mode)
The stop mode value changes the behavior of the measurement engine, after the Run key is
pressed.
• Hold: Measurement continues in the background.
• Stop: Measurement stops and runs again.
You can stop/hold the measurement using the Run key.