Factory Default Parameters - Dentsply Sirona X-Smart Plus User Manual

Table of Contents

Advertisement

Step-by-Step Instructions
When the desired continuous rotary file is selected, press the + or -
SPEED keys to select the desired speed setting.
When the speed value is changed from the default setting, SPEED is
displayed between brackets.
If the MEMO key is not pressed to save the setting, the setting will be
lost once a different file setting is selected.
The speed setting can be adjusted from 250 to 1000 rpm in increments
of 50 rpm and from 1000 to 1200 rpm in increments of 100 rpm.
Press the + or - TORQUE keys to select the desired torque setting.
When the torque value is changed from the default setting, TORQUE is
displayed between brackets.
If the MEMO key is not pressed to save the setting, the setting will be
lost once a different file setting is selected.
The torque setting can be adjusted from 0.6 to 4.0 Ncm in increments
of 0.1 Ncm.The preset torque and speed values of all continuous rotary
file systems can be altered individually.
CAUTION
Before using the motor handpiece, verify the correctness of the
changed parameters.
6.8

Factory Default Parameters

To return to the original default parameters, follow the general reset
instructions:
Turn the power off.
1.
Connect the AC adapter and check that the AC-IN lamp lights (see chapter
2.
6.5.1 Connecting the AC
Hold down the POWER key for more than 2 seconds, while pressing the
3.
MEMO key.
34/60
adapter).
BENXSPSDFUWEB / Rev.09 / 10-2018

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents