Keysight Technologies N4960A User Manual page 74

Serial bert 17 and 32 gb/s
Table of Contents

Advertisement

Operation
Label Name
OutEnable
Amp
Offs
Term
Coupling
Div
74
Refer to
Table 17
for the Div Clk Out menu descriptions.
Table 17. Div Clk Out menu descriptions
Description
Enables/disables the divided clock output.
Sets the amplitude of the divided clock. The optional units are mV and V. The
resolution is 0.005 V.
Sets the offset of the divided clock. The maximum range is a function of the
termination voltage setting. The optional units are mV and V. The resolution is
0.005 V.
The offset may be limited by the current termination voltage. If the desired
offset value conflicts with the current termination value, a message indicating
a conflict appears in the N4960A serial BERT controller display. Selecting Yes
to resolve the conflict will set the termination to an appropriate value.
Sets the termination voltage of the divided clock. The maximum range is a
function of the offset setting. The optional units are mV and V. The resolution
is 0.005 V.
The termination may be limited by the current offset voltage. If the termination
value conflicts with the offset value, a message indicating a conflict appears in
the N4960A serial BERT controller display. Selecting Yes to resolve the conflict
will set the offset to an appropriate value.
Selects AC or DC output coupling.
Coupling must be set to DC before offset or termination can be set to a non-
zero value. If an attempt to change the offset or termination to a non-zero
value conflicts with the AC coupling setting, a message indicating a conflict
appears in the N4960A serial BERT controller display. Selecting Yes to resolve
the conflict will set the coupling to DC.
Sets the divided clock divide ratio.
Keysight N4960A Serial BERT 17 and 32 Gb/s User Guide