Keysight Technologies N4960A User Manual page 172

Serial bert 17 and 32 gb/s
Table of Contents

Advertisement

Remote Operation
172
Command
:PG:DATA:LLEVel:AMPLitude?
Description
Return the amplitude value of the data logic level.
Example
:PG:DATA:LLEV:AMPL?
0.800V
Queries the amplitude of the pattern generator connected to channel 0
(Jitter) only.
Command
:PG:DATA:LLEVel:OFFSet
Description
Adjust the DC offset of the data eye presented at the data outputs from
–2.0 V to 2.0 V in 0.005 V increments. The units are V and mV. The
default value/unit is 0.0 V.
Example
:PG:DATA:LLEV:OFFS 2V (@ 1)
Sets the DC offset to 2 V for the pattern generator connected to channel
1 (Delay) only.
Keysight N4960A Serial BERT 17 and 32 Gb/s User Guide